7.3

DIO Interface

The DIO Interface provides 2 digital inputs which are hardwired to the Digital outputs. These may be used to detect
the state of external switches (output off) or to monitor the state of the outputs.

7.3.1 DIO Descriptor

Offset
0x00
0x01
0x02
0x03
0x04

7.3.1.1 DIO Sensor Type

The interface provides a bitmapped input of the 2 digital signal lines.

7.3.1.2 DIO Data Type/Format

7
Smart
Sensor
0
7.3.1.2.1 Data Type
7.3.1.2.2 Factory Calibrate
7.3.1.2.3 Sensor Value Writeable

7.3.1.3 DIO Input Configuration

7
Available
0
7.3.1.3.1 Lock
7.3.1.3.2 Apply Scaling
Name
Value
Sensor Type
0x18
Data Type/Format
0x46
Configuration
0x23
Sensor Device
0x0f
UOMR
"DIN"
Sensor Type
SI Derived Units
0x18
6
Sensor Value
Writeable
Calibrate
0
The 4-bit Data Type field determines the type of data of the specific sensor.
The Factory Calibrate bit is not used for DIO types.
This indicates that the sensor value may be overwritten. Not used on DIO inputs.
6
5
Assigned
Apply
Scaling
0
1
If set, the user-specified units of measure string (4 character maximum) will be used in place of
the default DIN.
If set, the user-defined Offset and Gain values will be used to adjust the sensor reading:
Digital Type (Bit mapped)
Configurable, Float type
Scaling applied, Bits 0 and 1 enabled
DIN bits enabled / inverted
Units of measure
Measurement
DIN
Bit mapped digital inputs
DIO Data Type/Format
5
4
Factory
reserved
0
0
DIO Input Configuration
4
3
Lock
?
Result = (Raw Reading * Gain) + Offset
Description
3
2
1
Data Type
6 == Floating point
2
1
Sub Channel Selection
0x03 == bits 0 and 1
0
0
33 |
M 5 7 5 1