Data synchronization is significantly slowed in two cases:
- When the display mode "Visualizing data" temperature is set (Picture 19: Home screen menu of
data )
- When the display mode "Detailed Data visualization of a capsule" (Picture 20: Detailed
visualization screen of a capsule) is set
In order to optimize data recovery, we recommend to leave the display mode.

6.7.7 Visualization of the end-of-life of the capsule

On the monitor, when a capsule reaches the end of life, the message "Low" appears in the
column Batt (Picture 20: Detailed visualization screen of a capsule) of the detailed visualization
screen.
The capsule will stop around 500 steps after the appearance of the first message "Low" (if the
monitor and the capsule are in continuous RF communication).

6.7.8 Min/Max reset

In order to reset the alarms and the Min / Max values shown in the data visualization screen,
go to the menu "DATA" and select the function "Min/Max reset".
These proceedings will involve all the capsules associated with the monitor.

6.7.9 Measurement period

It is possible to change the period between two temperature measurement. To perform the
modification, please go to pill menu and then Frequency configuration.
Picture 23: pill menu for frequency configuration
Picture 22: Measurement period
A list of five periods choice is proposed. By default the selected value is 30s. You need to be
carefull with this choice. Indeed, if you select 5s, you will be able to synchronise much quickly,
but the amount of data can increase dramatically. A large period will extend the autonomy of
the pill away from the monitor, but the synchronisation process could take more time.
Indeed, to synchronise data, the monitor has to be in real time RF communication with the pill
during several measurements. As soon as the RF communication is disrupted, the monitor
needs to wait for the next pill communication to continue synchronisation process. If the period
is short, the waiting time is shorter.