Requirements
Safety notes
Always switch the system off when performing work on the Memoskop.
Documentation required
This document RX57-029.061.01..applies to SIREMOBIL 2000, LITHOSTAR Multiline
and UROSKOP D1/D2.
MEMOSKOP
•
MEMOSKOP 2
•
MEMOSKOP 3/50/(100) and SUB 2000
SIREMOBIL 2000
•
Service instructions
LITHOSTAR Multiline
•
Setting the software parameters
•
Service software control
Uroskop D1/D2
•
Handling service software
•
Calling up the set-up mode (see page 5-1)
Measurement instruments, auxiliary devices and tools required
•
Digital multimeter Fluke 8060A
•
Service PC
•
Service tool kit
•
MEMOSKOP keyboard
•
Protective conductor meter
•
Serial PC connection cable for service PC
Protective conductor test
After finishing all work steps and attaching all covers, perform the protective conductor
test according to ARTD-002.731.16.
The protective conductor resistance must not exceed 0.2 O.
•
When working on parts connected to line voltage, there is the risk of electric shock.
•
To comply with EMC regulations, PC board shielding must be installed properly.
•
Always switch the system off or disconnect the power plug prior to removing the covers.