Documentation Required; Memoskop; Siremobil 2000; Lithostar Multiline - Siemens MEMOSKOP 2 SUB Service Instructions Manual

Table of Contents

Advertisement

Requirements
Safety notes
Always switch the system off when performing work on the Memoskop.

Documentation required

This document RX57-029.061.01..applies to SIREMOBIL 2000, LITHOSTAR Multiline
and UROSKOP D1/D2.

MEMOSKOP

MEMOSKOP 2
MEMOSKOP 3/50/(100) and SUB 2000

SIREMOBIL 2000

Service instructions

LITHOSTAR Multiline

Setting the software parameters
Service software control

Uroskop D1/D2

Handling service software
Calling up the set-up mode (see page 5-1)

Measurement instruments, auxiliary devices and tools required

Digital multimeter Fluke 8060A
Service PC
Service tool kit
MEMOSKOP keyboard
Protective conductor meter
Serial PC connection cable for service PC

Protective conductor test

After finishing all work steps and attaching all covers, perform the protective conductor
test according to ARTD-002.731.16.
The protective conductor resistance must not exceed 0.2 O.
When working on parts connected to line voltage, there is the risk of electric shock.
To comply with EMC regulations, PC board shielding must be installed properly.
Always switch the system off or disconnect the power plug prior to removing the covers.
Siemens AG
Medical Engineering
X2146
X2009
RR2-120.061.02..
RXL2-120.032.02..
RXL2-120.113.01..
RLL5-310.113.02..
97 02 101 Y4290
11 02 669 X2009
44 15 899 RV090
99 00 440 RE999
RX57-029.061.01
Rev. 02
10.97
1
Page 1 of 4 MEMOSKOP 2/3/50/100/SUB
TD SD 34
1 - 1
1
1
1
1
1
1
1