Electrical settings/programming

SIREMOBIL 2000

For the configuration of the MEMOSKOP 3/50/(100)
refer to chapter 5 of the service instructions RR2-120.061.02.. .
For the configuration of the MEMOSKOP SUB 2000
refer to chapter 5 of the service instructions RR2-120.061.02.. .
UROSKOP D1/D2
For the configuration of the Memoskop 50 / (100)
CAUTION
Switch the system "ON"
Wait for the MEMOSKOP to boot.
Activate set-up mode
NOTE
Actuate the following keys on the keyboard in sequence:
- <NEXT PAT>
- <@>
- <%>
- <CTRL> + <S>
After activating set-up mode, a sequence of numbers is displayed in the lower region of
the monitor image.
Enter the values indicated for the relevant TV system on the keyboard:
- for TV systems with 60Hz and 120 Hz memory output:
02 00 66 08 FE FF FF 01 02 32 63 63
- for TV systems with 50 Hz and 100 Hz memory output:
02 00 66 08 FF FF FF 01 02 32 63 63
Enter the numbers without any blanks.
Press the keys <CTRL> + <E>.
Exit set-up mode
Switch the system "OFF" and "ON" again.
The memory returns to normal operating mode.
Siemens AG
Medical Engineering
If PC board D18 on MEMOSKOP 2K or PC board D1 or D2 on
MEMOSKOP 50/(100) is replaced, a download of the MEMOSKOP
software must be performed on UROSKOP.
If PC board D1 is being replaced in MEMOSKOP 50/(100), the
MEMOSKOP must be configured additionally via the keyboard.
<x> means: press key x.
<CTRL> + <E> means: press and hold <CTRL> and simulta-
neously press key <E>.
RX57-029.061.01