Changing The Contact Time Tk; Changing The Postcontact Time Tn; Exit Setup - Siemens MEMOSKOP 2 SUB Service Instructions Manual

Table of Contents

Advertisement

7 - 2

Changing the contact time TK

Enter the following numbers: 00006508
**
means in HEX code: 01 = 100 ms, 02 = 200 ms, 32 = 5000 ms.
Then press the [CTRL] + [E] keys simultaneously.
Then, depending on the requirement, program the precontact time TV and/or the
postcontact time TN or exit the Hardcopy Setup (refer to the section "Exit Setup ").

Changing the postcontact time TN

Enter the following numbers: 0000650A
**
means in HEX code: 01 = 100 ms, 02 = 200 ms, 32 = 5000 ms
Then press the [CTRL] + [E] keys simultaneously.
Then, depending on the requirement, program the precontact time TV and/or the
contact time TK or exit the Hardcopy Setup (refer to the section "Exit Setup ").

Exit Setup

Press the [@] and [%] keys in sequence.
Switch the unit off and then back on.
After switching the unit back on, the new Hardcopy timing will be valid.
If plug D1.X6 was removed from the MEMOSKOP, reconnect it.
MEMOSKOP 2/3/50/100/SUB
Laser camera timing (not for MEMOSKOP 2)
RX57-029.061.01
Rev. 02
**
**
Page 2 of 4