Maintenance
Requirements
The requirements described in Chapter 1 of these instructions apply to maintenance as
well.

Visual inspection

Check the whole system for damage and secure mountings.

Electrical safety

Check that warning labels are complete and in good condition. Replace them, if
necessary.
Check cables and connectors for damage. Replace them, if necessary.
CAUTION

Cleaning

Clean the fan.
Remove any dust or dirt.

Functional inspection

Perform a functional check and image quality test in accordance with the associated
documentation (diagnostic test, IQ binder).

Final steps

With the system/unit closed, perform a protective conductor test.
Siemens AG
Medical Engineering
The protective conductor test is part of the safety inspection and
should be performed when maintenance work is complete and the
system is closed. Refer to "Final steps".