PC board replacement
Replacement of disk drive MS1
•
Open the MEMOSKOP.
•
Remove the large cover from MEMOSKOP.
•
Remove the disk drive.
•
Disconnect the plug.
•
Remove the metal holder from the drive.
CAUTION
•
Insert jumpers according to the table.
•
Attach the metal holder including spacers to the new drive.
•
Connect the plug to the new drive.
•
Insert the new drive in the location of the old one.
The drive must not contact any of the other components.
•
Reattach the cover.
•
Close the MEMOSKOP.
LPS, ELS, FB
-A0
A1
A2
*1)
SS
*1)
EP
WS
*1)
TE
*1)
PK
*1)
*2)
*3)
Siemens AG
Medical Engineering
Do not lose the spacers. They are required when installing the
disk drive to provide insulation.
Quantum
open
jumpered
jumpered
open
open
*1)
open
jumpered
open
All other existing jumpers remain open
If applicable
Substitute jumper, if applicable
Software VA07A must be loaded in the MEMOSKOP.
RX57-029.061.01
Rev. 03
Labeling
*3)
Seagate
ST 31 250 N
J4 Jumper
open
7-8
J4 Jumper
jumpered
9-10
J4 Jumper
jumpered
11-12
J1 Jumper
jumpered
1-2
J1 Jumper
jumpered
5-6
J2 Jumper
jumpered
5-6
10.97
Seagate
ST 51 080 N
J8 Jumper
5-6
J8 Jumper
3-4
J8 Jumper
1-2
J8 Jumper
23-24
J8 Jumper
*2)
28-30
J8 Jumper
*2)
32-34
Page 5 of 6
MEMOSKOP 2/3/50/100/SUB
TD SD 34
6 - 5
6
*3
open
jumpered
jumpered
jumpered
jumpered