For both environmental and lab calibrations,
wait at least five seconds for the sensor to
adjust to current sensing levels. Click the
Next button.
For lab calibrations, release the sensor from calibration meter. For environmental calibrations, take sensor
out of targeted environment.
For both environmental and lab calibrations,
wait at least five seconds for the sensor to
adjust to 0 sensing levels. Click the Finish
button.
*See Appendix B: Worksheets, pages 36
Page 26
Enter 0 in the lowest value text field and
click the Next button.
Note, that if performing an envi-
ronmental calibration at a frac-
tion of total sensor range, the
bar graphs in the main Global
Logger window will interpret
full scale range at maximum
calibration readings instead of
actual total sensor range. All
other logger operations will per-
form correctly. To revise cali-
bration settings so that the
graphs will function correctly,
follow the worksheet Calibrat-
ing the Bar Graph to Higher
or Lower Range in Appendix
B*.