Appendix
Electromagnetic immunity
The RadiForce series has been tested at the following compliance levels according to the
testing requirements for professional healthcare facility environments specified in IEC /
EN60601-1-2.
The customer or the user of the RadiForce series should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC / EN61000-4-2
Electrical fast
transients / bursts
IEC / EN61000-4-4
Surges
IEC / EN61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC / EN61000-4-11
Power frequency
magnetic fields
IEC / EN61000-4-8
34
Test level for
professional
healthcare facility
environments
±8 kV contact
±8 kV contact
discharge
discharge
±15 kV air discharge
±15 kV air discharge
±2 kV power lines
±2 kV power lines
±1 kV input / output
±1 kV input / output
lines
lines
±1 kV line to line
±1 kV line to line
±2 kV line to ground
±2 kV line to ground
0 % U
(100 % dip in
0 % U
T
U
) 0.5 cycles and 1
U
T
cycle
cycle
70 % U
(30 % dip in
70 % U
T
U
) 25 cycles / 50 Hz
U
T
0 % U
(100 % dip in
0 % U
T
U
) 250 cycles / 50
U
T
Hz
Hz
30 A/m
30 A/m
(50 / 60 Hz)
Compliance level
Electromagnetic environment -
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
(100 % dip in
Mains power quality should be
T
) 0.5 cycles and 1
that of a typical commercial or
T
hospital environment. If the user
of the RadiForce series requires
(30 % dip in
T
continued operation during power
) 25 cycles / 50 Hz
T
mains interruptions, it is
(100 % dip in
recommended that the
T
) 250 cycles / 50
RadiForce series be powered
T
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
The product should be kept at
least 15 cm away from the
source of power frequency
magnetic fields during use.
Guidance