Electronic Considerations; Power Cord; Usb Cord - Quansys Q-View IMAGER LS Operator's Manual

Table of Contents

Advertisement

ELECTRONIC CONSIDERATIONS

Power Cord

The power cord is a 3-wire grounding-type plug.
This cord will only fit into a grounded outlet, which
is a requirement for the Q-View Imager LS. Do not
remove the grounding leg of the plug. Doing so will
void all warranties on the Q-View Imager LS.
The power cord provided with the Q-View Imager LS
is specific to the power supplied in your region. If you
have a different voltage supply or need a different
power cord for any reason, please contact technical
support. Do not use an extension cord for the Q-View
Imager LS. Ensure that all equipment plugged into
the same circuit as the Q-View Imager LS does not
exceed the amperes of service, or you may experience
a tripped breaker or power failure that could result in
the loss of your image.
Before unplugging the power cord, make sure Q-View
Software is closed on the PC. Make sure the USB cord
is unplugged. At this point, it is safe to remove the
power cord.

USB Cord

The USB cord connects the Q-View Imager LS to the
PC, which controls and operates the Q-View Imager
LS. This cord should only be unplugged from the
Q-View Imager LS once Q-View Software is closed.