Focusing The Q-View Imager Ls; Light Leak Check - Quansys Q-View IMAGER LS Operator's Manual

Table of Contents

Advertisement

4
be running Q-View™ Software.
5. Open Q-View Imager LS lid, ensure the enclosure
where the glass/tray insert fits is free from
debris and liquids.
6. Clean the Q-View Imager
LS glass within the glass/
tray insert with a lint-
free cloth and ethanol,
rubbing
window cleaner.
7. Place the glass/tray insert
on the Q-View Imager
LS with the glass on the
bottom.
8. Open Q-View Software.
The connected imager should automatically be
detected and a dialog box should open with a
link to Manage Imagers. If this does not occur,
in the Settings menu, click Manage Imagers.
9. From the Manage Imagers menu the connected
imager may be selected, named, focused and a
dark field acquired.

Focusing the Q-View Imager LS

Open the Q-View Imager LS lid. Place the 96-well
focusing plate (included) on the plate tray. From the
Manage Imagers menu, select Focus. With the lid
still open, use the + and – buttons on the software
to bring the words printed on the focusing plate into
focus if needed. You can click on the image to zoom in
or out. When zoomed in, right-click and drag to scroll.

Light Leak Check

Upon completing the Q-View Imager LS installation
it may be necessary to ensure there are no light
leaks. To perform this test remove any plates from
the Q-View Imager LS, ensure the lid is closed, and
capture an image. The average pixel intensity of the
image should be less than 500. If the pixel intensity
is above this specification, please contact customer
support at Quansys Biosciences
(782-6797)
or support@quansysbio.com.
alcohol,
or
1-888-QUANSYS
For Research Use Only