Acquire Dark Field; Setup Completion; Disassembly; Moving The Q-View Imager Ls - Quansys Q-View IMAGER LS Operator's Manual

Table of Contents

Advertisement

Acquire Dark Field

Dark Field Image Acquisition minimizes background
noise in subsequent plate images. Remove any plates
from the Q-View Imager LS, close the Q-View Imager
LS lid, and click OK on the software. In the Settings
menu click on Manage Imagers. Select the correct
imager and press Acquire Dark Field. The process
will take approximately five minutes.

Setup Completion

Once the Q-View Imager LS setup is finished, the
software opens with the Update Preferences dialog
box. Select your preferences and log in if user accounts
were enabled. If your Internet connection on any
computer running Q-View Software goes through a
proxy, specify these settings under Administration
> Network Settings. Your Q-View Software and
Q-View Imager LS are now ready to use.

Disassembly

Carefully remove the glass/
tray insert from the Q-View
Imager LS. Take caution that,
while removing the glass/tray
insert,
nothing
is
dropped
into the Q-View Imager LS
cavity. Anything dropped will
fall directly onto the lens of
the camera and cause damage
which will adversely affect the
Q-View Imager LS.
MOVING THE Q-VIEW IMAGER
LS
Ensure there is no plate on the tray of the Q-View
Imager LS and unplug all cords. The Q-View Imager
LS weighs approximately 9.12 kg (20.12 lbs).
In order to move or ship the Q-View Imager LS, there
are several critical steps that should take place.
For Research Use Only
5