Diagnostics and servicing ....................43 Maintenance ........................43 Diagnosis function ......................43 Rectify faults ........................45 Exchange process module on version IP68 (25 bar) ............46 Instrument repair ......................47 Dismount..........................48 Dismounting steps......................48 WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 3
Dimensions ........................67 Trademark ........................75 Safety instructions for Ex areas Take note of the Ex specific safety instructions for Ex applications. These instructions are attached as documents to each instrument with Ex approval and are part of the operating instructions. Editing status: 2020-05-12 WIKA Operating Instructions - Process pressure transmitter IPT-2x...
The dot set in front indicates a list with no implied sequence. Sequence of actions Numbers set in front indicate successive steps in a procedure. Battery disposal This symbol indicates special information about the disposal of bat- teries and accumulators. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
During work on and with the device, the required personal protective equipment must always be worn. Appropriate use The IPT-2x is a pressure transmitter for process pressure and hydro- static level measurement. You can find detailed information about the area of application in chapter "Product description". Operational reliability is ensured only if the instrument is properly used according to the specifications in the operating instructions manual as well as possible supplementary instructions.
A Class 2 power supply unit has to be used for the installation in the USA and Canada. Exception: Versions with measuring ranges from 250 bar. These are subject of the EU Pressure Device Directive. Not fulfilled when connecting to an external display and adjustment unit. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Hardware from 1.0.0 • Software from 1.3.5 Note: You can find the hardware and software version of the instrument as follows: • On the type plate of the electronics module • In the adjustment menu under "Info" The type label contains the most important data for identification and Type label use of the instrument: WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Serial number of the instrument Symbol of the device protection class ID numbers, instrument documentation Principle of operation Application area IPT-2x is suitable for applications in virtually all industries. It is used for the measurement of the following pressure types. • Gauge pressure •...
Page 9
Fig. 3: Configuration of the measuring system with piezoresistive sensor ele- ment Sensor element Base element Transmission liquid Process diaphragm Strain gauge (DMS) sensor element For measuring ranges above 100 bar, a strain gauge (DMS) sensor element (dry system) is used. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 10
Absolute pressure: the measuring cell contains vacuum and is encapsulated. The ambient pressure is not compensated and does hence influence the measured value. Seal concept The measuring system is completely welded and thus sealed against the process. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
"Technical data", "Materials and weights". Supplementary cleaning procedures The IPT-2x is also available in the version "Oil, grease and silicone- free". These instruments have passed through a special cleaning procedure to remove oil, grease and paint-wetting impairment sub- stances (PWIS).
Page 12
3 Product description Lifting and carrying With instrument weights of more than 18 kg (39.68 lbs) suitable and approved equipment must be used for lifting and carrying. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
To maintain the housing protection, make sure that the housing lid is closed during operation and locked, if necessary. Devices with threaded fitting are screwed into the process fitting with Screwing in a suitable wrench via the hexagon. See chapter "Dimensions" for wrench size. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 14
Make sure that the upper temperature limits stated in chapter "Technical data" for the environment of the electronics hous- ing and connection cable are not exceeded. Fig. 6: Temperature ranges Process temperature Ambient temperature WIKA Operating Instructions - Process pressure transmitter IPT-2x...
When quickly opening/closing the housing cover, the measured value can change for approx. 5 s by up to 15 mbar. For an effective ventilation, the filter element must be always free from buildup. In case of horizontal mounting, turn the housing so that the filter element points downward after the instrument is installed. This provides better protection against buildup. Caution: Do not use a high-pressure cleaner. The filter element could be dam- aged, which would allow moisture into the housing. The following paragraphs describe how the filter element is arranged in the different instrument versions. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 16
This provides better protection against buildup. Fig. 8: Position of the filter element - Ex-d version Rotatable metal ring Filter element Instruments with absolute pressure have a blind plug mounted instead of the filter element. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Instruments with absolute pressure have a blind plug mounted instead of the filter element. Process pressure measurement Measurement setup in Keep the following in mind when setting up the measuring system: gases • Mount the instrument above the measuring point Possible condensation can then drain off into the process line. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 18
Siphon in U or circular form Pipeline A protective accumulation of water is formed through condensation in the pipe bends. Even in applications with hot steam, a medium temperature < 100 °C on the transmitter is ensured. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Mount the instrument below the min. level Do not mount the instrument close to the filling stream or emptying • area • Mount the instrument so that it is protected against pressure shocks from the stirrer Fig. 14: Measurement setup for the level measurement WIKA Operating Instructions - Process pressure transmitter IPT-2x...
In electroplating plants as well as plants for cathodic corrosion protec- tion it must be taken into account that significant potential differences exist. This can lead to unacceptably high currents in the cable screen if it is grounded at both ends. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
3. Loosen compression nut of the cable gland and remove blind plug 4. Remove approx. 10 cm (4 in) of the cable mantle, strip approx. 1 cm (0.4 in) of insulation from the ends of the individual wires WIKA Operating Instructions - Process pressure transmitter IPT-2x...
10. Reinsert the display and adjustment module, if one was installed 11. Screw the housing lid back on The electrical connection is finished. Single chamber housing The following illustration applies to the non-Ex as well as to the Ex-ia version. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Housing IP66/IP68 (1 bar) Wire assignment, con- nection cable Fig. 18: Wire assignment in permanently connected connection cable Brown (+) and blue (-) to power supply or to the processing system Shielding WIKA Operating Instructions - Process pressure transmitter IPT-2x...
5 Connecting to power supply External housing with version IP68 (25 bar) Overview Fig. 19: IPT-2x in IP68 version 25 bar with axial cable outlet, external housing Transmitter Connection cable External housing Electronics and connec- tion compartment for power supply 4...20mA...
• Internal check of the electronics • Indication of a status message on the display or PC • The output signal jumps to the set fault current WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 27
5 Connecting to power supply Then the actual measured value is output to the signal cable. The value takes into account settings that have already been carried out, e.g. default setting. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
5 s, the display returns to the main menu. The menu language is then switched over to "English". Approx. 60 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not confirmed with [OK] will not be saved. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
The return to the measured value indication is carried out through the [->] or [ESC] keys or automatically after 3 s Note: You can find a description of the individual steps in the quick setup guide of the sensor. You can find "Extended adjustment" in the next sub-chapter. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
The available digits include: • Letters from A … Z • Numbers from 0 … 9 • Special characters +, -, /, - WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 31
In this menu item you activate/deactivate the slave sensor for elec- tronic differential pressure and select the application. IPT-2x can be used for process pressure and level measurement. The setting in the delivery status is process pressure measurement. The mode can be changed in this adjustment menu.
Page 32
Adjustment IPT-2x always measures pressure independently of the process vari- able selected in the menu item "Application". To output the selected process variable correctly, an allocation of the output signal to 0 % and 100 % must be carried out (adjustment).
Page 33
[OK]. Span adjustment Proceed as follows: 1. Select with [->] the menu item Span adjustment and confirm with [OK]. 2. Edit the mbar value with [OK] and set the cursor to the requested position with [->]. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 34
[OK]. The cursor jumps now to the pressure value. 4. Enter the pressure value for the full vessel (e.g. 900 mbar) cor- responding to the percentage value. 5. Save settings with [OK] The max. adjustment is finished. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 35
The default setting is output characteristics 4 … 20 mA, fault mode < 3.6 mA. The device assumes an approximately constant temperature and static pressure and converts the differential pressure into the flow rate via the characteristic curve extracted by root. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 36
Russian • Italian • Dutch • Portuguese • Japanese • Chinese • Polish Czech • • Turkish In delivery status, the IPT-2x is set to English. In this menu item, you define which measured value is displayed. Display value 1 and 2 WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 37
In menu item "Peak value, temperature", both values are displayed. In another window you can carry out a reset of the two peak values separately. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 38
Any user programmable linearization curve as well as the measured value memory are deleted. The following table shows the default values of the instrument. De- pending on the instrument version or application, all menu items may not be available or some may be differently assigned: WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 39
Number of positions after the decimal point, automatically Backlight Switched on Diagnostics Menu item Parameter Default value Sensor status Peak value indicator Pressure Actual measured value Temperature Actual temperature values from measuring cell, elec- tronics Simulation Process pressure WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 40
The data are saved only after release. Scaling (1) In menu item "Scaling" you define the scaling variable and the scaling unit for the level value on the display, e.g. volume in l. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 41
Change the settings of the special parameters only after having con- tacted our service staff. 6.4.5 Info Device name In this menu item, you can read out the instrument name and the instrument serial number: WIKA Operating Instructions - Process pressure transmitter IPT-2x...
In the display and adjust- If the instrument is equipped with a display and adjustment module, ment module the parameter adjustment data can be saved therein. The procedure is described in menu item "Copy device settings". WIKA Operating Instructions - Process pressure transmitter IPT-2x...
General software error F105 The instrument is still in the start phase, Wait for the end of the switch-on phase the measured value could not yet be Measured value is deter- determined mined WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 44
Impermissible pressure value If necessary, use an instrument with a higher measuring range Tab. 7: Error codes and text messages, information on causes as well as corrective measures WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Check, adapt if necessary ance too high Current signal greater than Sensor electronics defective Replace device or send in for repair de- 22 mA, less than 3.6 mA pending on device version WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Proceed as follows when carrying out the exchange: 1. Losen the fixing screw with the hexagon key wrench 2. Carefully detach the cable assembly from the process module Fig. 25: IPT-2x in IP68 version, 25 bar and lateral cable outlet, external housing Process module Plug connector Fixing screw...
Complete one form for each instrument • If necessary, state a contamination • Clean the instrument and pack it damage-proof • Attach the completed form and possibly also a safety data sheet to the instrument WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. If you have no way to dispose of the old instrument properly, please contact us concerning return and disposal. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Transmission liquid with measuring ranges up to 40 bar. With measuring ranges from 100 bar dry measuring cell. Halocarbon oil: Generally in oxygen applications, not with vacuum measuring ranges, not with absolute meas- uring ranges < 1 bar WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 50
Ʋ Aluminium/Stainless steel housing 50 Nm (36.88 lbf ft) Glass with Aluminium and stainless steel precision casting housing Only for 316L with 3A approval Between transmitter and external electronics housing. Fix connected to the sensor. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 51
Nominal measuring ranges and overload capacity in psi Nominal range Overload capability Maximum pressure Minimum pressure Gauge pressure 0 … +5 psig +15 psig -14.5 psig Data on overload capability apply for reference temperature. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 54
Max. output current 21.5 mA Load See load resistance under Power supply Damping (63 % of the input variable), 0 … 999 s adjustable Dynamic behaviour output Dynamic characteristics depending on medium and temperature WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 55
Deviation (according to IEC 60770-1) Applies to the digital signal output (HART, Profibus PA, Foundation Fieldbus) as well as to the ana- logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio "nominal measuring range/set span". WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 56
(factor FMZ) and Turn Down (factor FTD). The additional factors are listed in the following tables. Additional factor through accuracy class Accuracy class 0.075 %, 0.1 % 0.2 % Factor FMZ Additional factor through Turn Down Different availability depending on measuring range and process fitting WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 57
TD 20 : 1 Factor FTD 1.75 10.5 Thermal change current output through ambient temperature Applies also to the analogue 4 … 20 mA current output and refers to the set span. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 58
-20 … +60 °C (-4 … +140 °F) -20 … +60 °C (-4 … +140 °F) ble PE With ceramic/metallic measuring cell with gold-coated diaphragm, the values must be multiplied with factor 3. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 59
40 °C (104 °F) 0 °C (32 °F) 85 °C 105 °C (185 °F) (221 °F) Fig. 30: Temperature derating IPT-2x, version up to +105 °C (+221 °F) Process temperature Ambient temperature Process fittings acc. to DIN 3852-A, EN 837 WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 60
(104 °F) 0 °C (32 °F) 85 °C 150 °C (185 °F) (302 °F) Fig. 31: Temperature derating IPT-2x, version up to +150 °C (+302 °F) Process temperature Ambient temperature SIP process temperature (SIP = Sterilization in place) Vapour stratification for 2 h +150 °C (+302 °F) Process pressure see specification "Process pressure"...
Page 61
(140 °F) 0 °C (32 °F) 110 °C 150 °C (230 °F) (302 °F) Fig. 32: Temperature derating IPT-2x, version up to +150 °C (+302 °F) Process temperature Ambient temperature 80 °C (176 °F) 65 °C (149 °F) 0 °C (32 °F)
Page 62
Ʋ Standard length 5 m (16.40 ft) 2 g with housing version stainless steel double chamber IP66/IP68 (0.2 bar), only with absolute pressure. Breather capillaries not with Ex-d version. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 63
< 14 V) Ʋ for U ≤ 1.0 V (16 … 400 Hz) 24 V DC (18 V < U < 35 V) Load resistor Ʋ Calculation )/0.022 A Ʋ Example - with U (24 V - 9.6 V)/0.022 A = 655 Ω = 24 V DC WIKA Operating Instructions - Process pressure transmitter IPT-2x...
With a 4 … 20 mA output, the thermal change of the current output F must be added: = √((F + (F + (F perf Galvanic separation between electronics and metal housing parts When used with fulfilled housing protection. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Tab. 28: Determination of the additional factor measuring cell for above example: F Turn Down TD 1 : 1 TD 2.5 : 1 TD 5 : 1 TD 10 : 1 TD 20 : 1 Factor FTD 1.75 10.5 WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 66
The total deviation of the measurement is hence 0.57 %. Deviation in bar: 0.57 % of 4 bar = 0.023 mbar The example shows that the measurement error in practice can be considerably higher than the basic deviation. Reasons are temperature influence and Turn down. WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Fig. 36: Housing versions in protection IP66/IP67 and IP66/IP68 (0.2 bar) Plastic single chamber (IP66/IP67) Aluminium - single chamber Stainless steel single chamber (electropolished) Stainless steel single chamber (precision casting) Stainless steel single chamber (electropolished) IP69K WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 68
(2.76") 8 mm (0.12") (0.32") 110 mm x 90 mm (4.33" x 3.54") Fig. 37: IP68 version with external housing Lateral cable outlet Cable outlet axial Plastic housing Stainless steel housing, electropolished WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 69
G½ A inside G¼ (ISO 228-1) ½ NPT, inside ¼ NPT (ASME B1.20.1) ½ NPT PN 1000 For the version with "Second Line of Defense", the measure of length increases by 17 mm (0.67 in). WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 70
3 and 4 with temperature adapter and screen sheet for 180 °C/200 °C 1½ NPT (ASME B1.20.1) For the version with "Second Line of Defense", the measure of length increases by 17 mm (0.67 in). WIKA Operating Instructions - Process pressure transmitter IPT-2x...
Page 71
(3.05") ø 105 mm (3.60") (4.13") Fig. 40: IPT-2x, hygienic fitting 150 °C (piezoresistive/strain gauge measuring cell) Clamp 2" PN16 (ø64mm) DIN 32676, ISO 2852 Hygienic fitting with compression nut F 40 PN 25 Varivent N 50-40 PN 25 Collar socket DN 40 PN 40, DIN 11851...
Page 72
ø 78 mm (2.60") (3.07") (3.07") Fig. 41: IPT-2x, hygienic fitting 150 °C (metallic/ceramic measuring cell) Clamp 2" PN16 (ø64mm) DIN 32676, ISO 2852 Hygienic fitting with compression nut F 40 PN 25 Varivent N 50-40 PN 25 Collar socket DN 40 PN 40, DIN 11851...
Page 73
0.94" 6“ 4xø 0.75" 5“ 0.06" Fig. 42: IPT-2x, flange connection 150 °C (piezoresistive/strain gauge measuring cell) Flange connection according to DIN 2501 Flange connection according to ASME B16.5 Order-specific Order-specific For the version with "Second Line of Defense", the measure of length increases by 17 mm (0.67 in).
Page 74
6“ 4xø 0.75" 5“ 0.12" Fig. 43: IPT-2x, flange connection 180 °C/200 °C (ceramic/metallic measuring cell) Flange connection according to DIN 2501 Flange connection according to ASME B16.5 Temperature adapter up to 180 °C Temperature screen sheet up to 200 °C...
9 Supplement Trademark All the brands as well as trade and company names used are property of their lawful proprietor/ originator. WIKA Operating Instructions - Process pressure transmitter IPT-2x...