Glomax ® 20/20 Luminometer Fluorescent Module; Installing The Fluorescent Module - Promega GloMax 20/20 Technical Manual

Luminometer
Table of Contents

Advertisement

9.
GloMax
20/20 Luminometer Fluorescent Module
®
The Fluorescent Module (Cat.# E5351 and E5361) enables measurement and analysis of common fluorophores. The
following Fluorescent Modules are available for different excitation and emission wavelengths:
Blue: excitation, 460nm; emission, 515–575nm
UV: excitation, 365–395nm; emission, 440–470nm
The type of Fluorescent Module required for a particular assay depends on the excitation and emission wavelengths of
the fluorophore or dye. The Fluorescent Module is designed for easy installation and removal. Store the Fluorescent
Module in the product case when not in use. Keep liquids, dust and debris away from direct contact with the Fluorescent
Module.
9.A. Installing the Fluorescent Module
1.
Turn off the GloMax
2.
Align the metal pins of the Fluorescent Module with the module port inside the luminometer (Figure 9).
3.
Slide the Fluorescent Module forward and press down to lock it into place.
4.
Turn on the GloMax
5.
If this is the first-time use of the Fluorescent Module, you must identify the type of module. Touch the button
that corresponds to the name of the module. If you are not sure of the identification of your module, contact
Promega Technical Services.
Figure 9. Aligning the Fluorescent Module with the module port.
12
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM276 · Revised 5/21
20/20 Luminometer.
®
20/20 Luminometer.
®
Module Port
Fluorescent Module
Light Standard