Battery Life - Fluke DATAPAQ XDL12 User Manual

For furnace tracker
Table of Contents

Advertisement

Logger Status LEDs
Red
5 flashes, alternating
with green LED
Flashing, alternating with
green LED, at sample
interval *
On
Flashing together with
green LED, at sample
interval *
Off
Flashes 5 times (once
per second)
Flashing every second
Flashing every 5
seconds
2 quick flashes every
second
*
Flashing interval will actually fall in range 1–5 s.
Battery
The logger uses a pack of three rechargeable 1.2 V nickel-metal-hydride
(NiMH) cells. Only battery packs supplied by Datapaq are suitable.
Service life of the rechargeable NiMH battery is about three years or 1,000
charge/discharge cycles. The logger must be returned to Datapaq for battery
replacement.
For subsidiary power, the logger also contains a lithium AA battery. This will
normally last for the life of the logger, but, if replacement is needed, the logger
must be returned to Datapaq.

Battery Life

Battery life (discharge time) of NiMH rechargeable batteries is affected by the
following factors.
DATAPAQ XDL12
Green
5 flashes, alternating with
red LED
Flashing, alternating with red
LED, at sample interval *
Flashing at sample interval *
Flashing together with red
LED, at sample interval *
Flashing at sample interval *
Off
Off
Off
Off
Meaning
Logger successfully reset
Logger awaiting trigger
Logger awaiting trigger but one or
more thermocouples is open circuit
All probes are above trigger
temperature, and thus data-recording
cannot be triggered by rising
temperature (or, if falling trigger is set,
all probes are below trigger point)
Logger acquiring data
Connection between communications
lead (attached to an active PC) and
logger has been made
Internal error
Logger has data in memory which has
not been downloaded
Logger too hot to start logging (after
pressing Start button)
Specifications and Operation