Checking The Data - Fluke DATAPAQ XDL12 User Manual

For furnace tracker
Table of Contents

Advertisement

Troubleshooting
Logger Communications Problems
Communications lead not fully inserted – Check correct sockets are
being used.
Damaged communications lead or connectors – Check for breaks
and other damage; replace the lead.
Batteries not charged – Recharge the batteries, ensuring the charging
LEDs are illuminated (see p. 12).
To minimize communications problems:
Connect the lead first to the PC and then to the logger.
Always use the same USB port – the one which was first used to set
up communications (p. 21).
Logger Download Error Messages
Error Message
There are
insufficient readings
in the logger
Logger stopped due
to low battery
Logger memory full Data collection may have stopped before the run was completed:

Checking the Data

Thermocouple probes are generally reliable, but damage resulting from
inappropriate use or handling can produce erroneous readings. If you suspect
that invalid data may have been introduced into your temperature profile
(paqfile), select the View Data tab in the Insight software's Analysis Window to
view the raw data as downloaded from the logger. The various types of invalid
data which may be contained in a paqfile are shown in the analysis grid as
follows.
DATAPAQ XDL12
Action
Check trigger set point (time or temperature).
Check logger's battery for charge.
Check date/time settings on computer.
Check probes and their connections.
Reset logger and test probes (see 'Logger Diagnostics', below).
Replace or recharge the battery as appropriate, then repeat the
profile run.
check the data collection period and sample interval before resetting
the logger for another run (see 'Resetting the Data Logger', p. 23).
Troubleshooting
33