Fluke DATAPAQ XDL12 User Manual page 11

For furnace tracker
Table of Contents

Advertisement

Data storage
Data-collection start
Pre-trigger data stored
Hot-data protection
Communications
PC/software compatibility
Logger reset
Battery
Battery life
Battery charger
Specifications for Specific Thermocouple Types
Type J
Normal
0°C to 800°C
range
32°F to 1,472°F
Accuracy *
±0.3°C
±0.5°F
Lower range
Accuracy *
Resolution
0.1°C/0.2°F
Socket color
Black
(IEC 584)
Type R
Normal
50°C to 1,760°C
range
122°F to 3,200°F
Accuracy *
±1.1°C
±2.0°F
Lower range
−40°C to 50°C
−40°F to 122°F
Accuracy *
±1.8°C
±3.2°F
Resolution
0.1°C/0.2°F
Socket color
Orange
(IEC 584)
*
Using sample interval > 0.8 s over temperature range shown. For more details, contact Datapaq.
DATAPAQ XDL12
18,000 data-points per channel
Start button, Temperature rise, Temperature fall, Date/time
Yes (configurable; see p. 28)
By non-volatile memory, and software warning if reset attempted
before download
USB
See p. 19
Possible after download by start button (using previous run
parameters)
Datapaq NiMH rechargeable battery-pack + lithium AA
Over 60 hrs continuous use (see p. 13), depending on sample
interval and operating temperature
All regions except Americas – CH0070
Americas – CH0075
Type K
−100°C to 1,370°C
−180°F to 2,498°F
±0.3°C
±0.5°F
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
0.1°C/0.2°F
Green
Type S
50°C to 1,760°C
122°F to 3,200°F
±1.1°C
±2.0°F
−40°C to 50°C
−40°F to 122°F
±1.7°C
±3.1°F
0.1°C/0.2°F
Orange
Specifications and Operation
Type N
−100°C to 1,300°C
−180°F to 2,372°F
±0.4°C
±0.7°F
−190°C to −100°C
−310°F to −180°F
±0.5°C
±0.9°F
0.1°C/0.2°F
Pink
Type T
−100°C to 400°C
−180°F to 752°F
±0.3°C
±0.5°F
−196°C to −100°C
−321°F to −180°F
±0.5°C
±0.9°F
0.1°C/0.2°F
Brown
11

Advertisement

Table of Contents
loading

Table of Contents