RF300 – Help with Restart or Factory Reset
Issue 3 – 03/01/2017
This help document refers to the RF300 Diligence WiFi Data Loggers, including the
RF324 Alarm Unit.
How do I Restart or Reset my RF300 Diligence WiFi device?
Under normal circumstances, you do not need to reset your RF300 Diligence WiFi Device. However, you may be
instructed to do this by our support team to recover a Device which is not working properly, or return it to its
factory state.

Restart

Performing a Restart on the WiFi Sensor will restart the internal firmware, retaining all configuration and
network settings. The Sensor will disconnect from the wireless network, and immediately enter a low power
sleep mode. It can take up to 30 minutes for the Sensor to re-connect and synchronize after a Restart. Under
normal circumstances the Sensor will synchronize within 15 minutes. Once communication is re-established the
Sensor returns to its normal transmission rate. A Restart will cause a new recording session to be started.

To perform a Restart...

Comark Instruments
52 Hurricane Way, Norwich, Norfolk, NR6 6JB, UK
Tel: +44 207 942 0712
www.comarkinstruments.com
Comark Instruments
PO Box 500, Beaverton, OR 97077, USA
Tel (503) 643 5204
www.comarkUSA.com
RF313-TH
From the Sensor Home screen
Press and hold the button for 10 seconds
Release when the LOW symbol starts to flash.
The Professionals' Choice

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Comark RF300 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Fluke Comark RF300 Series

  • Page 1 RF300 – Help with Restart or Factory Reset Issue 3 – 03/01/2017 This help document refers to the RF300 Diligence WiFi Data Loggers, including the RF324 Alarm Unit. How do I Restart or Reset my RF300 Diligence WiFi device? Under normal circumstances, you do not need to reset your RF300 Diligence WiFi Device. However, you may be instructed to do this by our support team to recover a Device which is not working properly, or return it to its factory state.
  • Page 2 All other Sensor types RF311-T, RF313-TP, RF314-TC From the Sensor Home screen Press and hold the button for 10 seconds Release when the LCD shows rSt RF324 (Alarm) Press and hold the button for 10 seconds Release when the LEDs start to cycle. Factory Reset Performing a Factory Reset on the WiFi Sensor will clear all internal configuration and network settings.
  • Page 3 All other Sensor types RF311-T, RF313-TP, RF314-TC From the Sensor Home screen Press and hold the button for 20 seconds Release when the LCD shows FArS RF324 (Alarm) Press and hold the button for 20 seconds. The LEDs will cycle. When the LEDs go out, release the button.

This manual is also suitable for:

Comark rf313-thComark rf311-tComark rf313-tpComark rf314-tcComark rf324

Table of Contents