Testing and
Manufacturer's Recommendations
Operating
Requirements
Ambient
Operating Temperature, Humidity and Altitude
Condition
Restrictions for
Operation
IMPORTANT NOTES:
Quality

SECTION III: Quality Control

Controls
Overview
Internal Controls:
prior to testing.
operating system.
PASS/FAIL QC
Results
Controls Run at
Start-up
AUGUST 3, 2021
The SQA-iO is sensitive to vibrations during the testing cycle. Operate away from
devices that cause electronic noise or vibrations such as centrifuges.
Un-plug the device when not in use for an extended period of time.
Ambient temperature limitations: The SQA-iO is calibrated to conduct tests at room
temperature: 20-25ºC (68-77ºF). This is the recommended range for room
temperature maintenance of semen samples prior to testing for up to 1 hour following
collection.
Sample processing limitations: The device is calibrated to test semen samples at
room temperature. Heating samples in ANY WAY will impact results. Do not heat
samples.
Semen is considered biologically hazardous material and is subject to
individual laboratory protocols for handling such materials. At a minimum,
it is recommended to:
Wear laboratory coat, mask and gloves when handling semen
o
Dispose of samples post testing in designated hazardous waste containers
o
Require that only personnel trained to work with biologically hazardous
o
materials test and handle semen samples.
Maximum operational humidity is up to 80% for temperatures of up to 31
decreasing linearly to 50% humidity at 38
Operates in a wide range of ambient temperatures (15-38
calibrated
to
20-25
º
C (68-77
º
F).
Intended for indoor use at a maximum altitude of 2000m, mains supply fluctuations
±10%, Overvoltage Category II, Pollution Degree II.
Humidiy exceeing the limitations above may impact test results due to condensation
on the optical detectors. Assess ambient humidity and temperature prior to device
operation.
Ambient temperatures exceeding the limitation above may impact the accuracy of
semen motility test results because of the known effect of temperature on human
semen.
A series of tests are automatically run when the SQA-iO is turned on and
These internal QC tests check the calibration settings and the internal
Internal Controls are run @ SQA-iO Start-up: PASS/FAIL results are reported
on the SQA-iO home screen along with troubleshooting information
Stabilization and autocalibration:
checked to ensure they are in proper range by analyzing the system sensors. Once
stable for 30 seconds the device passes stabilization and autocalibration. A warning
message is displayed on the home screen if there is a failure.
System noise:
Measures the electronic noise level of the system to insure effective
measurement of electronic signals.
Self-test:
Electronic signals simulating motility and concentration measurements
verify that the calibration settings are consistent with factory specifications.
SQA-iO Service Manual
º
C.
measure
semen
samples
System stability and reference parameters are
6
º
C with
º
C) however the system is
at
room
temperature: