Service Manual Version 187.5.1
Motility Channel
Light disturbances caused by moving sperm cells are converted into analog signals.
Technology
Proprietary algorithms interpret the signals and report motility.
Concentration Channel
Based on spectrophotometry analysis of the semen samples and application of
proprietary algorithms.
1. The capillary is inserted into the measurement compartment.
2. Sample concentration is evaluated in the "tall" 10 mm chamber of the capillary.
3. Motility is detected in the "thin" 0.3 mm section of the capillary.
4. This information is then digitized and routed to the microprocessor that applies

SECTION II: System Specifications and Requirements

SQA-iO Device
Overview and
System
Components
Minimum requirements:
Front Panel: Measurement Compartment (Capillary insertion for testing)
Measurement
Compartment
AUGUST 3, 2021
SQA-iO Service Manual
Concentration
detector
#1
Concentration
LED
algorithms to extract the required clinical semen parameters.
Dimensions: 8 X 9.5 X 10.5 cm
Weight: 0.350 Kg
Power supply: 5V DC (USB)
Noise level: 0 [dBA]
Device power consumption: 1.7 [BTU/hour] = 0.5 [Watts]
PC: Intel Core i5 M520 2.4GHz or equivalent
RAM: 4GB
Monitor Screen: Color, Wide screen – minimum resolution 1024 x 768
Operating system compatibility: Windows 7 Professional x32 or above
Communication Ports: one USB port
Internet Access: 5mb per second
Capillary
chamber
Motility
detector
Analog to
#3
Digital
Converter
#2
Motility
LED
Operational
Monitor
4
Self-Test
and QC
#4
Semen
parameters
extraction
algorithms
KeyPad