4.4

Fixing the Base

4.4.1
Concrete Floor
To fix the base on the concrete floor:
Required Tool
Parts and
Accessory
1.
Mark the base's location on the floor.
Green X / X18 (Model: PHT-75CHS) Installation Manual
4. Installing the X-ray Unit: Base Stand Type
L = 200 mm
Hammer Drill
Tips: 3 mm - 8 mm
Ratchet Wrench
(0.12" - 0.3")
Hammer
5/16 x 60 w/ Spring
and Flat Washers -
Anchor Bolt
(Part No. 34)
Spring washer
M8 - 2 pcs
(7+-.9")
N/A
2 pcs
55