2. Pre-Installation Requirement
2.2

Electrical Installation Requirement

These specifications are based on the MEIGaN (Medical electrical installation
guidance notes).
Consult the companion manual for further information: Volume 3: Specification for
Electrical Installation.
2.3

Electrical Requirement

12
This equipment must be connected to a grounded outlet to
fulfill the safety provisions specified in IEC 60364: the 2
edition (2006).
Both PC and equipment must use the same power line if
connected to an MPSO.
Whenever not possible, use different power outlets for each
device. If a multiple portable socket outlet (MPSO) must be
used, ensure that the PC and equipment are connected to
the same MPSO.
Use a dedicated power outlet for the power cord. Failure to
do so may result in unstable system operations caused by
power fluctuations.
It is strongly recommended that you install an AVR. An AVR
(automatic voltage regulator) maintains a constant voltage
and allows for continuous operation during power
fluctuation.
Green X / X18 (Model: PHT-75CHS) Installation Manual
nd