High Intensity LED Stroboscope Digital Tachometer
DT-361/365
Instruction manual
Be sure to read before use.
Before use, please carefully read these safety precautions as well as instructions,
and follow them for proper use.
GlobalTestSupply
www.
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

Summary of Contents for Nidec SHIMPO DT-361

  • Page 1 All manuals and user guides at all-guides.com 98585A High Intensity LED Stroboscope Digital Tachometer DT-361/365 Instruction manual Be sure to read before use. Before use, please carefully read these safety precautions as well as instructions, and follow them for proper use. GlobalTestSupply www.
  • Page 2 All manuals and user guides at all-guides.com Safety Requirements Be sure to observe Before operation, maintenance and inspection, please carefully read this instruction manual and follow it for proper use. Please carefully read all information related to this unit's safety precautions before use.
  • Page 3: Table Of Contents

    All manuals and user guides at all-guides.com - Contents - Overview Before use Checking accessories Peel off the protection sheet Charging (only for DT-365) 2.3.1 Charging method 2.3.2 Indication of low battery voltage 2.3.3 Battery replacement Part names and functions Main unit Operation section Display...
  • Page 4: Overview

    All manuals and user guides at all-guides.com Overview A stroboscope tachometer is a measurement instrument to measure the speed (cycle) of rotating objects that rotate at a constant speed, or moving objects that repeatedly operate at a constant cycle. When the rotation (motion) cycle matches the flash cycle while the strobe flash is periodically applied on a rotating or moving object, the rotating (moving) object image appears to stand still.
  • Page 5: Peel Off The Protection Sheet

    All manuals and user guides at all-guides.com Peel off the protection sheet Peel off the protection sheet on the operation section. Operation section Peel off the (transparent) protection sheet Charging (only for DT-365) Be sure to charge the battery before initial use. Before charging, be sure to check that the power is turned OFF.
  • Page 6: Indication Of Low Battery Voltage

    All manuals and user guides at all-guides.com 2.3.2 Indication of low battery voltage If the remaining battery power drops to near zero, the number display alternates between the process setting and LLLLLL. However, operation can be performed even when the battery is in this low condition.
  • Page 7: Part Names And Functions

    All manuals and user guides at all-guides.com Part names and functions Main unit Super luminosity LED (18 lights) Operation section External I/O connector Power switch * The image is for DT-365. Operation section ⑤  ⑥  ⑦  ④  ③ ② ⑧  ① ...
  • Page 8: Display

    All manuals and user guides at all-guides.com Display 3.3.1 Part names Charging lamp (only for DT-365) Memory number display Numerical display Mode display Unit display Emission setting display 3.3.2 Number display •Indicates the setting value of the emission count (frequency) in internal oscillation emission mode. •Indicates the external signal frequency in external synchronous emission mode.
  • Page 9: Functions And Operations

    All manuals and user guides at all-guides.com Functions and operations Power ON/OFF Press the power switch when the power is OFF to turn the power ON. When the power is turned ON, the model is indicated, followed by internal oscillation emission or external synchronous emission.
  • Page 10: Emission Setting

    All manuals and user guides at all-guides.com Emission setting Each press of the SIG key switches between internal oscillation emission and external synchronous emission. Operation Display Internal oscillation emission External synchronous emission Each press switches the emission settings GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
  • Page 11: Internal Oscillation Emission

    All manuals and user guides at all-guides.com Internal oscillation emission 4.3.1 Display of internal oscillation emission Setting value Mode display Memory number display Display showing internal oscillation emission 4.3.2 Switching the modes Each press of MODE key switches the setting display value to FPM (emission count mode), DUR (emission duration mode), and PHA (phase mode).
  • Page 12: Emission Count Mode

    All manuals and user guides at all-guides.com 4.3.4 Emission count mode The emission count (frequency) can be set in the emission count mode (FPM). 4.3.4.1 Emission count setting Turn the dial to the CW direction to increase the emission count (frequency), and to the CCW direction to decrease it.
  • Page 13: Changing The Emission Count (Frequency) To Double Or Half

    All manuals and user guides at all-guides.com 4.3.4.2 Changing the emission count (frequency) to double or half The emission count (frequency) can be changed to double or half of the current setting value using the key operation on internal oscillation emission. 1) To change the emission count (frequency) to double of the current setting value Each press of the x2 key changes the emission count (frequency) to double of the current setting value on internal oscillation emission.
  • Page 14: Emission Duration Mod

    All manuals and user guides at all-guides.com 4.3.5 Emission duration mode The emission time (ratio) can be set by 0.1º within the range between 0.1º/ 360º and 3.6º/ 360º in the emission duration mode (DUR). The emission time setting value shows the strobe emitting time by angle while a rotating object rotates one revolution (360º).
  • Page 15: Phase Mode

    All manuals and user guides at all-guides.com 4.3.6 Phase mode When the rotation (motion) cycle of a measured object matches with the strobe flash cycle, the measured object appears to stand still. To change the angle (position) to make the object stand still, use the phase mode.
  • Page 16: External Synchronous Emission

    All manuals and user guides at all-guides.com External synchronous emission External synchronous emission is the function to emit a strobe flash in synchronization with an external pulse. •To activate external synchronous emission mode, refer to 4.2 Emission setting. • You can set which edge of the external pulse triggers the emission, the rising edge or falling edge. Refer to 4.6.3 Trigger edge setting.
  • Page 17: Emission Duration Mode

    All manuals and user guides at all-guides.com 4.4.2 Emission duration mode The emission time (ratio) can be set by 0.1º within the range between 0.1º/ 360º and 3.6º/ 360º in the emission duration mode (DUR). (The emission time setting value shows the strobe emitting time by angle while a rotating object rotates one revolution (360º).) While the “UNIT”...
  • Page 18: Phase Mode

    All manuals and user guides at all-guides.com 4.4.3 Phase mode Delay emission can be set within the input signal range between 60 fpm and 10,000 fpm. In the phase mode (PHA), the delay angle from the external pulse entry to strobe flash emission can be changed by 0.1º...
  • Page 19: How To Set The Delay

    All manuals and user guides at all-guides.com 4.4.3.1 How to set the delay The delay angle from the pulse input to emission can be set by 0.1º within the range between 1º and 359º by setting one cycle of the external pulse to 360º. Measure the external pulse frequency per cycle, and calculate the delay angle based on the measured frequency to perform the delay emission.
  • Page 20: Memory Function

    All manuals and user guides at all-guides.com Memory function Each setting value can be saved and read out. 4.5.1 Saving each setting value This function is used to save each setting value. There are two types of saving methods, one is to save a parameter by pressing and holding the “LOAD(SAVE)”...
  • Page 21: Saving On Internal Oscillation Emission

    All manuals and user guides at all-guides.com 4.5.1.1 Saving on internal oscillation emission Press and hold the “LOAD(SAVE)” key in the Internal oscillation emission mode to save the setting value. Press and hold the LOAD(SAVE) key to enter the save standby status. SAVE and the memory number in which the setting value is saved are indicated alternately.
  • Page 22: Saving On External Synchronous Emission

    All manuals and user guides at all-guides.com 4.5.1.2 Saving on external synchronous emission Press and hold the “LOAD(SAVE)” key in the external synchronous emission mode to save the setting value. Press and hold the LOAD(SAVE) key to enter the save standby status. SAVE and the memory number in which the setting value is saved are indicated alternately.
  • Page 23: Reading Out Each Setting Value

    All manuals and user guides at all-guides.com 4.5.2 Reading out each setting value This function is used to read out each setting value. 4.5.2.1 Reading out on internal oscillation emission Press the “LOAD(SAVE)” key in the internal oscillation emission mode to read out the setting value. Each press of the “LOAD(SAVE)”...
  • Page 24: Reading Out On External Synchronous Emission

    All manuals and user guides at all-guides.com 4.5.2.2 Reading out on external synchronous emission Press the “LOAD(SAVE)” key in the external synchronous emission mode to read out the setting value. Each press of the “LOAD(SAVE)” key changes the memory number to call. The memory number being read is indicated on the memory number display.
  • Page 25: Function Mode

    All manuals and user guides at all-guides.com Function mode The following settings in the table can be configured in the function mode. Setting item Default setting value Function mode 1 Measurement range setting 120,000 Function mode 2 Trigger edge setting Function mode 3 Auto emission stop time setting Function mode 4...
  • Page 26: Measurement Range Setting (Function Mode 1)

    All manuals and user guides at all-guides.com 4.6.2 Measurement range setting (Function mode 1) This function is used to set the measurement range on internal oscillation emission. The measurement range has two levels. The indicated contents depend on the measurement unit setting.
  • Page 27: Trigger Edge Setting (Function Mode 2)

    All manuals and user guides at all-guides.com 4.6.3 Trigger edge setting (Function mode 2) In the external trigger mode in which the edge of the external pulse triggers emission, the rising edge or falling edge can be set. When the trigger edge is set to L-H External pulse (Pulse entered from an external device to DT-361/365)
  • Page 28 All manuals and user guides at all-guides.com Turn the dial to the CW direction to set the trigger edge to H-L (falling). Turn the dial to the CCW direction to set the trigger edge to H-L (rising). Operate the dial to stop the alternate display and only the trigger edge setting value is indicated. Operation Display Indicates alternately...
  • Page 29: Auto Emission Stop Time Setting (Function Mode 3)

    All manuals and user guides at all-guides.com 4.6.4 Auto emission stop time setting (Function mode 3) This setting is used to stop the emission automatically when no operation is performed for a certain period of time. The available setting time is within the range between 0 min and 120 min, which can be changed by 1 minute.
  • Page 30: Input Signal Setting (Function Mode 4)

    All manuals and user guides at all-guides.com 4.6.5 Input signal setting (Function mode 4) Signals to be entered for external synchronous emission can be selected either from open collector input or voltage pulse input. Set it to ON to select open collector input, and OFF to select voltage pulse input. For details about the circuit, refer to 4.7.2 External pulse input.
  • Page 31: External I/O Connector

    All manuals and user guides at all-guides.com External I/O connector 4.7.1 Specifications for external I/O connector and pin assignment •DT-361 Pin number Signal name Remark ─ ─ DC 12 V power output External pulse output External pulse input External pulse input/output common ─...
  • Page 32: External Pulse Input

    All manuals and user guides at all-guides.com 4.7.2 External pulse input Connect the unit to external devices (sensors, etc.) to allow the strobe to emit light using the pulse signal from those devices in the external trigger mode. (For details, refer to 4.4 External synchronous emission) Available input frequency : Available measurement range 40 to 35,000fpm Available delay emission range 60 to 10,000fpm...
  • Page 33: Specifications

    All manuals and user guides at all-guides.com Specifications Model DT-361 DT-365 Application AC power input model Charging battery built-in model Emission count 60 to 120,000fpm Setting accuracy ±0.02% Available to set to the range between 60 fpm and 120,000 fpm, or 60 fpm Measurement range setting and 12,000 fpm (Set in function mode 1)
  • Page 34: External Dimensions

    All manuals and user guides at all-guides.com •Saving the setting value when the power is OFF Memory function • Saving 5 setting values each on internal oscillation and external synchronous emission respectively Available to set continuous emission, or set the time to stop emission by 1 minute within the range between 1 min and 120 min. Auto emission stop (Set in function mode 3) Built-in NiMH battery...
  • Page 35: Trouble Shooting

    All manuals and user guides at all-guides.com Trouble shooting Symptoms Factors Causes Treatment DT-365 does not flash and the Battery voltage is low Battery is not charged Please charge the battery display shows LLLLLLL. Battery lifetime Replace the battery The display of DT-365 shows Something wrong with Less capacity in the Please send it for repair if the...