TNP Digital Torque Meter
Operation Manual
Do not attempt to repair or alter this instrument. Warranty will be voided
and damage to the unit may result.
Use and store within the stated temperature and humidity ranges, or
damage and failure may result.
Ensure during testing unit is properly used on a stable, level surface.
Do not test products that are filled with liquid as the TNP torque tester is
not protected from liquid spills.
Only use electrical charger supplied with unit.
The Series TNP Digital Torque Meter is a precision instrument for
measuring OPEN and CLOSE torque values for various applica-
tions. The TNP Torque Meter is used in medical, pharmaceutical,
industrial quality control, R&D and countless other torque appli-
cations.
As a cap torque analyzer, the TNP repeatedly determines bottle
closure and breakaway torque. Its wide memory capacity is ca-
pable of storing 1,000 data points, including Average, Maximum
and Minimum values. A HI-GO-LO LED comparator feature can be
utilized for large batch testing to ensure overall product quality.
The TNP is a portable device powered with internal rechargeable
batteries or through provided universal AC adapter/charger.
Designed with a small footprint, it can easily be transferred be-
tween the shop floor and laboratory taking up little space when
operating. The TNP Torque Meter comes standard with USB con-
nection and cable for easy PC/Software interface. Combined with
the free "Digitorq" Software, statistical testing, graphing and anal-
ysis can be performed.
Each package includes the following:
• Operations Manual
• Software Installation Disk (Digitorq Software)
• Warranty Card
• USB A to USB B communication cable
• Universal AC adapter (100-240 VAC)
• Set of 30 mm chuck pins (4 pieces)
• TNP unit
Find Quality Products Online at:
SPECIFICATIONS
Measuring Range: TNP-0.5: 0-50.00 Ncm, 0-500.0 mNm,
0-5099 gcm, 0-4.425 lbin; TNP-2: 0-2.000 Nm, 0-200.0 Ncm,
0-20.39 kgcm, 0-17.70 lbin; TNP-5: 0-5.00 Nm, 0-500.0 Ncm,
0-50.99 kgcm, 0-44.25 lbin; TNP-10: 0-10.00 Nm, 0-1000 Ncm,
0-102.0 kgcm, 0-88.5 lbin
Test Sample Diameter Range: 0.39" - 7.48" (10 - 190 mm)
Accuracy: ±0.5% / F.S.
Display: Main display: 4-digit LCD display, Character height 0.5"
(12 mm)
Comparator Display: HI - GO - LO (green and red LED indica-
tor)
Display Update Time: Select from 0.125 second (8 times/sec-
ond), 0.25 second (4 times/second), 0.5 second (2 times/second),
1 second (1 time/second). Peak display: 0.125 second
Measuring Mode: Opening Mode: Show max value when open-
ing (peak display) Displays max counterclockwise torque; Closing
Mode: Show max value when closing (peak display) Displays max
clockwise torque; Average Mode: Real time display (average value
display) Displays torque in real time
Overload: 150% of full scale
Sampling Rate: 1000 Hz
Memory: 1000 data (max)
Data Output: USB1.1
Power: Built in nickel hydride battery (NiMH) or Auto-ranging AC
adapter (AC100 - 240V)
Battery Information: NiMH: 8 hours use after full charge. 16
hour max. recharge time.
Operating Environment: 32 - 104°F (0 - 40°C)
Dimensions: 12.60" x 8.94" x 4.96" (320 x 227x 126 mm)
Product Weight: TNP-0.5, -2, -5 = Approx. 17.6 lb (8 kg); TNP-
10 = 27.5 lb (12.5 kg)
Package Weight: TNP-0.5, -2, -5 = Approx. 19 lb (8.6 kg); TNP-
10 = 30.7 lb (13.9 kg)
Software: DigiTorque TNP Free Download from www.shimpoinst.
com
Included Accessories: USB cable, 120/240 VAC AC adapter,
30mm chuck pin (4pcs)
Certifications: CE
Warranty: 2 year
GlobalTestSupply
www.
.com
sales@GlobalTestSupply.com

Advertisement

loading

Summary of Contents for Nidec TNP Series

  • Page 1 TNP Digital Torque Meter Operation Manual Do not attempt to repair or alter this instrument. Warranty will be voided and damage to the unit may result. Use and store within the stated temperature and humidity ranges, or damage and failure may result. Ensure during testing unit is properly used on a stable, level surface.
  • Page 2 Parts and Functions Measuring mode indicators Power key Recall Key Mode Key Memory Key Zero/RST Key Main Display Comparator LED indicators Parts Functions/Description Turns the unit ON and OFF. Power key Recall data stored in OPEN and CLOSE modes. Recall key NOTE: in Average or Mean Mode, the Recall button has no function Memory key for storing data in Open and Close Modes.
  • Page 3 Main Display (Description) Units/Status Indicator Battery Indicator Main Display/ Setup Value Direction Indicator Extended Description • Units/Status indicator – located on the upper part of the display this shows the current units of measure selected for the torque meter. It also serves as a sub display under function mode (F01, F02, F03, etc.) Example: “OVR”...
  • Page 4 • Battery Indicator – shows the status of the TNP internal battery. “LO BAT” appears on the display to indicate a low battery status. “BAT” is shown when the TNP is charging. This indicator disappears when the battery is fully charged or when the AC adapter is disconnected from the TNP.
  • Page 5 If the display shows the normal operation of the torque meter, and no F01 is seen on the display, the Zero/RST key was released earlier than what is required. Turn off the torque meter and repeat the process. Function Mode Table FUNCTION Sub display Options...
  • Page 6 Operation diagram of how to access the Function Mode Turn OFF the TNP before starting Press and hold the Zero/RST key Zero/RST Power on then release the Power button Power but continue to hold Zero/RST key until display F01 Function Mode – Measurement unit TNP-2, 5, 10 Unit: N-m Unit: N-cm...
  • Page 7 Function Mode – Display (update rate) Mode Mode 1time/sec 2times/sec 4times/sec Mode 8times/sec Mode Zero/RST Pressing Zero/RST, it moves to the next function setting. Function Mode – Auto power OFF Set auto power OFF to Disable auto power OFF 10 minutes Mode Mode Zero/RST...
  • Page 8 Function Mode – Upper comparator value (Hi Limit) Present upper value Recall Recall Plus 1 by pressing Recall key Move to the right Move to the right Move to right side one by one by pressing Mem key. Press Mem key scrolls back to Recall the 1 digit position,...
  • Page 9 Operation Pretest Checklist: Determine the size of the sample to be tested. Adjust the 4 jig on the testing table accordingly. Center the sample and use the knob to secure the sample in place (clockwise to tighten the jig to the sample material). It is important to make sure that all the jigs are flush to the moving brackets.
  • Page 10: Comparator Function

    Mean Close Open Mean Close Open Zero/RST Resetting the Peak values to Zero MEAN is real time value based on the average data captured in 1000 samples per second. The maximum display update for all modes is 8 times/second. This update rate can be adjusted by changing F02 from function mode (See section - Values available 1, 2, 4, 8 times/second).
  • Page 11 How to change the HI and LO values from the TNP torque meter? • To set the HI and LO limits from the TNP torque meter access the function mode. (See Diagram How to Access the Function). • Power off the torque meter •...
  • Page 12 Zero adjustment / Tare the TNP Taring or zeroing the value of the TNP initialize the torque meter to zero. This function is performed by pressing Zero/RST button on the front panel. In OPEN and CLOSE modes this zeros out the Peak values measured. It is essential that the TNP be zeroed out before performing another test.
  • Page 13 How to recall stored Memory Select Open or Close Mode, press the Recall key and the display will indicate the number of data stored in memory. Press Recall key second time to access the following information. • AVERAGE Value • MAX Value •...
  • Page 14 Picture above indicates the memory ID followed by the stored data. Pressing the mode button anytime exits out of the memory window. Below is a flowchart of how the recall function works. Recall Average value Maximum value Power key Recall Recall Recall Last memory number, data display...
  • Page 15 Clearing Stored Data There are two types of memory clear available on the TNP torque meter. • Single memory clear • Clear all memory Single Memory Clear Single clear refers to erasing the stored data manually from the torque tester. The erase process starts from the most recent to the very first data stored.
  • Page 16 Here is a flowchart of how the single clear works. Clear Last memory number Zero/RST (Appear 1 sec.) When there is one last data, the last data is cleared by pressing Zero/RST key. After that, “nonE” is displayed at main display for 2 seconds, then the display returns to measuring mode as shown below: Last memory number No data...