93591B
DIGTAL TORQUE METER
TNX
SERIES
Instruction Manual
Please read the following content before operation
Please read instruction manual and [safty notes]carefully before
operation,the operate correctly.
www.nidec-shimpo.co.jp

Advertisement

loading

Summary of Contents for Nidec TNX Series

  • Page 1 93591B DIGTAL TORQUE METER SERIES Instruction Manual Please read the following content before operation Please read instruction manual and [safty notes]carefully before operation,the operate correctly. www.nidec-shimpo.co.jp...
  • Page 2: Safty Notes

    Safty notes please comply Please read instruction manual carefully before installation ,operation,maintenance,check,read ,product knowledge,safty information,not before operation. [Danger],[warning],[cantion]differentiate the importance of safty notes in the instruction manual,they are very import.Please comply. Danger mark indicates o possible death,serious injury or fire if the user Danger disregards the instruction .
  • Page 3 Cauction Don't charge with other instuction except Don't use except AC 100v AC connect instrution ,it may leads to circuit It may lead to fire and electrical shock fire . damage and fire. Please insert power cable. into AC connet Don't plug into or unplug from AC connect instrument while use in the loose condition,it instrument.
  • Page 4 1.Characteristic of this product ● Because it suit for P.P,so you can land save function 1st ,2nd ,3rd ,close,each of which are 1000 details. ● Can test the torque of open.close direction. ● Can save graduate land value to USB storge. ●...
  • Page 5 3.2 Operation panel ⑧ Test indicate LED ① POWER KEY ② RECALL KEY ④ Open/CloseKEY ⑤ FILE KEY ③ MEM KEY ⑥ ZERO/RST KEY ⑦ Lcd indicate part ⑨ Compare test determination indicate LED ① POWER KEY Use for ON/OFF the power ②...
  • Page 6 3.3.2 Express unit Express unit . Indicate"OVR"when exceed load Indicat"PWR"one minute before turn off the power . N・m Unit N・cm Unit At the overload Auto power off before 1 minute Blinking the LED of Open,Mean,Close 3.3.3 Charging display Indicate fllowing contents accrding to charge state. LO BAT Blinking :...
  • Page 7 5.1 Open test form Test the max torque presured on test table of open direction.     Display cycle  :  8 times per second.     Resert max height point by press ZERO/RST key. ZERO Open Mean Close Open Mean Close The opening directions ZERO +...
  • Page 8: Measuring Method

    6.Function of test instrument ・ Compare determinant upper/lower limit value with test data,max or store data loaded.LED light and indicate result. ・ Can set determinant upper/lower limit value of 1st,2nd,3rd separately.(reher to set operation of determinant upper/lower limit value of P12.1) ・...
  • Page 9 8.storage instrument function 8.1 1 store land When in open/close test mode,through keep pressing MEM key to land the max value to storge open test mode 【open test mode】 after one second,resert the max value and return to test mode. In case of exceeding 1000 data.
  • Page 10 【Close test mode】 It returns to original in 1 second. When data exceed 1000 Memory record After 1second,peak resert 8.2 Memory indication Memory indicate open test mode,then press RECALL key to enter memory indication mode. Memory data of memory indication mode are finally indicated. 【When open 1st/close choosing】...
  • Page 11 【When open 2nd/3nd choosing】 testing RECALL Cancel memory Enter final memory code indicate data ZERO Indicate memory code RECALL ・     ・ RECALL RECALL Enter final memory code indicate data Indicate memory data After indicating for one second After indicating for 1 second After indicating for 1 second After indicating for one second when F01=2nd...
  • Page 12 8.3 Cancel final data When final code,deta indicating keep pressing ZERO/RST key to cancel Data of final memory code. Enter final memory code indicate data Indicate clear Enter final memory dat indicate data ZERO After indicating for 1 second After indicating for 1 second when F01=2nd After indicating for 1 second After indicating for 1 second...
  • Page 13 9.USB memory save 9.1Ways for saving USB memory USB memory save is to save .Peal hold value inputed in memory and it statistical data by keep pressing FILE key when in open/close test mode. Besides,when saving in USB memory,determinating upper/lower limit by determinant upper /lower limit value setted and save the results. Accessoral USB memory of approriative connector for USB memory on right side of mainframe close test mode.
  • Page 14 9.3 Memory Data File Format oper(3rd)Memory Data Open                 Date,Jul/18/2007 15:29:24 Model,TNX-5          Unit,Nm                    (4) Data,10                    (5) Step,3rd                   (6) Denominator for SD,N-1   (7) ,1st,2nd,3rd Maximum,0.427,0.367,0.429   (8) Minimum,0,0,0   (9) Average,0.2047,0.1646,0.1839    (10) Standard Deviation,0.1281,0.1187,0.1188     (11) Difference,187.8,216.3,193.7    (12) Upper Limit,0.1,0.2,0.3   ...
  • Page 15 Denominator of standard deviation,variation,and fraction defective are calculated as follows. ● In case of N-1 ∑ ( Xi -ave.value) ave.value= ∑Xi/N denominator of standard= =measuring data N=memory data number ● n case of N ∑ ( Xi -ave.value) denominator of standard= ave.value = ∑Xi/N =measuring data N=memory data number...
  • Page 16: Function Setting

    11.Function setting 11.1 Function mode Function modes are as follows Item Sub display Setting conterds Initial setting Open peak torque 1st、 2nd、 3rd change Measuring unit N ・ m、 N ・ cm change N ・ m Display cycle 1、 2、 4、 8minites change A/Dfelter 3、...
  • Page 17: Auto Power Off Setting

    < Display cycle setting > In measuring mode display cycle is selectable from 1 time/second,2times/second,4 times/second,or 8 times/second. ※ Cnrrent setting is displayed first 1 time/second 2 time/second 4 time/second Open Open Close Close Open Close 8 time/second Open Close ZERO A/D filter setting <...
  • Page 18 < Communication device sitting > Set it connect voith printer or compnter ※ Cnrrent setting is displayed first Connecting with computer Connecting with printer Open Close Open Close Setting of outside output ZERO < Setting of outside output > Set the single of ouerweighed ortput,outside output or ortside output of compare instrument. ※...
  • Page 19 12. Set determinant upper/lower limit value 12.1 Setting operation of determinant upper/lower limit value. Set daterminant upper/lower limit value on 4 lines that no singles,then decide decimal point.According to unit and type.    Choose 0,2,3,4...9 according through pressing recau key on 1~4 lines(press relall key to return o when 9) move right by pressing mem key Expless the location of choosing unit decimae point Set the determinate lower limit value of compare instrument by pressing o key.
  • Page 20 Please set throngh press the key as set determinant upper limit value of 1st. return to the setting of determinant upper limit value as check out abnormality in power of in_out. When F01=1sth,set towards to deterninant upper limit value of close. 2nd ...
  • Page 21 13. Acquirement of data 13.1 USB commuricating function Connect the senior function digital tester to computer with accessorial USB wires,then communicating data with computer. (Please install the tied software on the leftside of lomputer.) 13.2 Characteristic of digitorg_TNX Bbecans of digitorg TNX ,advertising sofware of exced,you can get the test data of senier function data and store data from excel directly.
  • Page 22 14. Convenient funltion 14.1 load the file set Digitorq_TNX can make file ard load from Upanel whele save to TNX. Set file indude message of deterninant upper/lower linmit value,peak torque of open/close and(open/close)test mode. It can operate outmatically when comect with power sourle.If panel Uis conmected with mairframe.And if these is memery archives of test mode in set file,the archives will sewe in panel U automatically ard then concd.
  • Page 23 15.Operation of time setting In condition of power OFF FILE + POWER ※ Please input last two digits of Christian Era.   It can set up from 2007 to 2106. ※ Cnrrent setting is displayed first <Years settings> RECALL Press recall key Press recall key Year indications and1...
  • Page 24 16. Extrnal linker ・ Cable (optional) 2m only for a printer in length Pin number Signal names 16.1 Distribution of pin ① N/A ※ 1 ② N/A ※ 1 RxD (RS-232Creceive data) ③ printer → mauirframe ④ Digital GND ⑤ testing connection TxD (RS-232CTransmission data ⑥...
  • Page 25 17.Exceptant It will lead to short cirucit if connected with other tool,even arises five. Please don't charge with other tool beyond accessorial AC adapter When the customer arrive near the ferronickel battery,it will discharge possibly.so please charge by AC adapter. Connect HC adapter with its inward linker.insert socket of AC 100v.
  • Page 26: Specification

    19.Specification Model TNX-2 TNX-5 TNX-10 Measuring torque 2N ・ m 5N ・ m 10N ・ m Measuring range 0.000 ~± 2.000N ・ m 0.000 ~± 5.000N ・ m 0.00 ~± 10.00N ・ m Measuring unit N ・ m,N ・ cm exchange 0.000 ~±...
  • Page 27 MEMO...
  • Page 28 :1 Terada Kotari, Nagaokakyo-city, Kyoto, 617-0833 Japan Phone: 81-75-958-3608 FAX:81-75-958-3647...