Dt-200lr series handheld digital laser tachometer (4 pages)
Summary of Contents for Nidec Shimpo FG-L1
Page 1
Loadmeter Model FG-L1 v1.0 For use with Shimpo force gauge models: FGV-0.5, FGV-1, FGV-2, FGV-5, FGV-10, FGV-20, FGV-50, FGV-100, FGV-200H, FGV-300H, FGV-500H Congratulations on your purchase of Shimpo’s Loadmeter software. We trust you will enjoy many years of professional results from this product. Please read this entire instruction manual thoroughly before initial set-up and operation;...
TABLE OF CONTENTS Features and Benefits ....................... 3 Important Safety Precautions ................... 3 Software Installation and Cable Connection..............3 Requirements for Software Installation and Use ............3 Connecting an FGV Series Force Gauge to Your PC..........3 Software Installation Instructions ................3 Loadmeter Main Screen Description ................
Features and Benefits Shimpo’s Loadmeter software offers many features and benefits, including: • Allows user to display measurement results in a graph of force vs. time • Displays and records last, maximum, minimum, average, and standard deviation values (for spc applications) •...
Tool Bar Icons Opens file Saves file Prints recorded data Selects COM port Selects trigger Selects time interval (x-axis) Selects force (y-axis) Selects y-axis gram/kilogram force units Selects y-axis Newton units Selects y-axis lb. units Starts recording Stops recording Increases force interval Decreases force interval Tare function adjusts nonzero force value to zero Obtains measurement data at specified time interval...
Main Menu Display Displays Trigger Mode Displays Current Status of Loadmeter Displays Trigger Level Value Displays Trigger Level Direction Displays Maximum Peak Value Displays Maximum Measured Value Displays Trigger Position Displays Minimum Measured Value Displays Time and Date of Measurement Displays Total Number of X-axis Increments Displays Current Y-axis Force Measurement Displays Comment, i.e.
[Open] 1. Select [Open] as described above in “[File] Menu”. 2. Select appropriate drive and folder of file you wish to open, if not already selected. 3. Open the highlighted file by clicking [OK]. [Save] The Loadmeter software allows saving files in two formats.
[Print] Selecting [Print] as described in “[File] Menu” on page 6 prints data and a graph of recorded measurement. See example below. [Exit] Selecting [Exit] as described in “[File] Menu” on page 6 exits Loadmeter software. Page 8 of...
[Condition] Menu [Trigger] Sets trigger force to begin measurement [Time] Sets x-axis measurement increments, 0.5 second to 10 minutes [Weight] Sets y-axis measurement increments (g/kg, N or lb.) [COM Port…] Selects COM Port for input from force gauge [Trigger] [Mode] [Direction [Range] [Load Level]...
[Time] 1. Select [Time] as described in [Condition] File on page 9. 2. Select time interval, x-axis. Interval may be selected in 0.5-sec, 1-sec, 2-sec, 5-sec, 10-sec, 20- sec, 0.5-min, 1-min, 2-min, 5-min or 10-min. [Weight] 1. Select [Weight] as described above in [Condition] Menu on page 9.
Table Data Display Selecting the Table icon as described in “Tool Bar Icons” on page 5 and setting trigger to [None] as described in “[Condition] Menu” on page 9 will record force and time when the Catch icon is selected. Selecting Table icon as described in “Tool Bar Icons”...
Troubleshooting The following are general checkpoints; please call your local Shimpo representative or contact Shimpo Instruments at (800) 237-7079 for further assistance. Can not find "Setup" file to install Loadmeter software. Double-click the [My Computer] icon on the Windows desktop, double-click [CD-ROM drive], double-click the Loadmeter folder, then double-click the folder named “Disk1”.
Need help?
Do you have a question about the Shimpo FG-L1 and is the answer not in the manual?