Safety notes Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
2 x 1.5 V AAA Weight 100 g / <1 lb Total dimensions PCE-PH 26F 185 mm x Ø 40 mm / 7.3 x 1.6 in PCE-PH 28L 240 mm x Ø 40 mm / 9.44 x 1.6 in Electrode dimensions PCE-PH 26F 40 mm x Ø...
Delivery scope PCE-PH 26F 1 x pH meter PCE-PH 26F 1 x buffer 4.01 pH powder for 250 ml solution 1 x buffer 7.00 pH powder for 250 ml solution 1 x buffer 10.01 pH powder for 250 ml solution 2 x batteries 1.5 V AAA...
Keys Function - Switches the meter ON/OFF - Freezes the reading. Press the key again to continue MEAS / Hold measurement. - Ends the calibration or adjustment and returns to measuring mode - Press the key to start the calibration. - Press and hold the key to open the setup menu.
Switch meter on / off To switch on the meter, press and hold the "MEAS" key until the display switches on. Press and hold the "MEAS" key for 5 seconds to switch it off. The auto-off function switches the unit off after 8 minutes without a key being pressed. The function can be enabled or disabled in the basic settings.
MENU FACTORY DESCRIPTION OPTIONS ITEM SETTING (pH4.01/7.00/10.01) Selection of the calibration points NIST (pH4.01/6.86/9.18) 1 point Number of calibration points 2 points 2 points 3 points Celsius Selection of the temperature °C unit Fahrenheit Temperature calibration Enables the hold function when the measured value is stable The auto-off function switches...
1-point calibration 6.1.1 The 1-point calibration must be selected in the basic settings. 6.1.2 Rinse the electrode with distilled water. Press the "CAL" key. Depending on the setting, the display shows "pH 7.00/CAL1" or "pH 6.86/CAL1". 6.1.3 Immerse the electrode in the pH 7.00 (or pH 6.86) buffer solution with the end of the electrode completely immersed in the calibration solution.
2-point calibration 6.2.1 The 2-point calibration must be selected in the basic settings. 6.2.2. Carry out the calibration of the first calibration point as described under 6.1.2. … 6.1.4. The display shows "CAL 2". 6.2.3 Rinse the electrode with distilled water. Immerse the electrode in the pH 4.01 or pH 10.01 (pH 9.18) buffer solution with the end of the electrode completely immersed in the...
3-point calibration 6.3.1 The 3-point calibration must be selected in the basic settings. 6.3.2. Carry out the calibration of the first calibration point as described under 6.1.2. … 6.1.4. The display shows "pH 4.01" and "CAL 2". 6.3.3 Rinse the electrode with distilled water Immerse the electrode in the pH 4.01 buffer solution with the end of the electrode completely immersed in the calibration solution.
6.3.7. press the "ENTER" key. The meter starts the calibration. 6.3.8 After the measured value has stabilised, the display shows the slope of the electrode in % and then displays "End". The 3-point calibration is completed. Press the "MEAS" key to return to measuring mode even during calibration. Temperature calibration When deviating temperature values are displayed, you can adjust them in the corresponding menu item.
pH measurement Check the electrode as described in chapter 5.3 and switch on the meter, see chapter 5.4. Rinse the electrode with distilled water. Immerse the electrode in the medium with the end of the electrode completely immersed in the medium.
Warranty You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms. Disposal For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
Page 15
PCE Instruments contact information Germany France Spain PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel.