Agilent Technologies 1100 Series User Manual page 20

Lc devices connecting the hardware and triggering data acquisition
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Configuring the Autosampler for Contact Closure
12-digit hexadecimal number
located on the back of the
HP JetDirect network card
12
Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition
• In the Stack IP Address box, type the IP address for your Agilent 1100 LC stack.
Leave the value set to the default or contact your network administrator for the IP
address.
The default stack IP address is 172-16-0-102.
• In the Sub-mask box, type the subnet mask (address mask). Leave the subnet mask set
to its default value, or contact your network administrator for the subnet mask.
The default subnet mask value is 255-255-255-0.
• In the Bootp Server box, type the IP address for the network card in your PC that is
responsible for assigning the stack IP address for your Agilent 1100 LC system. Leave
the value set to the default or contact your network administrator for the BOOTP
server IP address.
The default IP address for the network card is 172-16-0-101.
6. Click OK to save the settings and close the dialog box.
Figure 6.
Agilent1100 Autosampler dialog box
7. Click Done to exit the Xcalibur Instrument Configuration program.
8. Open the Xcalibur data system before you turn on the devices of your Agilent1100 Series
LC stack.