Agilent Technologies 1100 Series User Manual page 14

Lc devices connecting the hardware and triggering data acquisition
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Connecting the Hardware
Installing a Network Card in an Agilent Module
6
Agilent 1100 Series LC Devices Connecting the Hardware and Triggering Data Acquisition
2. To prevent damage to the interface board, ensure that you are using ESD protection. For
more information on preventing damage caused by an electric discharge, refer to the
Reference Manual for the Agilent module.
CAUTION To prevent damage to an instrument, always use ESD protection when
handling electronic boards and components.
3. To remove the cover plate from the slot where you plan to install the network interface
card, loosen the two screws that fasten the plate to the chassis of the module.
shows the location of the cover plate on the back panel of an Agilent 1100 Series detector.
Figure 2.
Back panel of an Agilent 1100 Series detector
ANALOG
ANALOG
RS232
4. Record the MAC address (12-digit hexadecimal number) located on the back of the HP
JetDirect 400N network card. You will need this number to complete the configuration
of the Agilent stack. See
5. Insert the HP JetDirect 400N network card into the slot, and then tighten the two screws
that fasten the PCB to the chassis of the module.
Cover plate
CAN
RS232
CAN
Figure 6
on
page
12.