Table of Contents

Advertisement

Quick Links

Operating instruction
PIOX S72*
UMPIOX_S72xV1-9EN

Advertisement

Table of Contents
loading

Summary of Contents for Flexim PIOX S72 Series

  • Page 1 Operating instruction PIOX S72* UMPIOX_S72xV1-9EN...
  • Page 2 PIOX is a registered trademark of FLEXIM GmbH. FLEXIM GmbH Boxberger Straße 4 12681 Berlin Germany Tel.: +49 (30) 936 67 660 Fax: +49 (30) 936 67 680 E-mail: info@flexim.com www.flexim.com Operating instruction for PIOX S72* UMPIOX_S72xV1-9EN, 2022-05-15 Article number: 25841 Copyright (©) FLEXIM GmbH 2022...
  • Page 3: Table Of Contents

    PIOX S72* Table of contents Table of contents Introduction ................7 Safety instructions .
  • Page 4 Table of contents PIOX S72* Measurement ................100 10.1 Enabling the analysis .
  • Page 5 PIOX S72* Table of contents Data transmission ..............170 19.1 Service interfaces .
  • Page 6 Table of contents PIOX S72* 2022-05-15, UMPIOX_S72xV1-9EN...
  • Page 7: Introduction

    Copyright The contents of this operating instruction are subject to changes without prior notice. All rights reserved. No part of this operating instruction may be reproduced in any form without FLEXIM's written permission. UMPIOX_S72xV1-9EN, 2022-05-15...
  • Page 8: Safety Instructions

    The measuring equipment has to be checked for proper condition and operational safety before each use. If troubles or damages have occurred during installation or operation of the measuring equipment, please inform FLEXIM. It is not allowed to make unauthorized modifications or alterations to the measuring equipment.
  • Page 9: Not Intended Use

    • With the exceptions stated in chapter 12, the measuring equipment is maintenance-free. Any components and spare parts may only be replaced by FLEXIM. The operator shall carry out periodic checks for changes or damages that can present a danger. For further information, contact FLEXIM.
  • Page 10: Safety Instructions For Transport

    → Wear the required personal protective equipment. → Observe the applicable rules. • If you detect a transport damage when unpacking the delivery, please contact the supplier or FLEXIM immediately. • The transmitter is a sensitive electronic measuring instrument. Avoid shocks or impacts.
  • Page 11: General Principles - Analysis

    For the standard data sets, the physical quantities for analysis from Tab. 3.2 are selectable. Customized data sets with special physical quantities for analysis can be created in the FLEXIM laboratory. UMPIOX_S72xV1-9EN, 2022-05-15...
  • Page 12: Terms

    3 General principles – analysis 3.2 Terms PIOX S72* Terms Tab. 3.1: Terms term explanation analyte components of a mixture which are to be chemically analyzed matrix components of a mixture which are not to be analyzed The transmitter can be used for the determination of the following physical quantities for analysis: Tab.
  • Page 13: Basics - Flow Measurement

    4 Basics – flow measurement PIOX S72* 4.1 Measurement principle Basics – flow measurement In the ultrasonic flow measurement, the flow velocity of the fluid in a pipe is determined. Further physical quantities are derived from the flow velocity and from additional physical quantities, if necessary. Measurement principle The flow velocity of the fluid is measured using the transit time difference correlation principle.
  • Page 14 4 Basics – flow measurement 4.1 Measurement principle PIOX S72* 4.1.2 Measurement of the flow velocity The signals are emitted and received by two transducers alternatively in and against the flow direction. If the fluid is flowing, the signals propagating in the fluid are displaced with the flow. Caused by this displacement, the sound path of the signal is reduced in flow direction and increased in the opposite direction.
  • Page 15: Measurement Arrangements

    4 Basics – flow measurement PIOX S72* 4.2 Measurement arrangements Fig. 4.3: Transit time difference Δt Δt 1 – signal in the flow direction 2 – signal against the flow direction Measurement arrangements 4.2.1 Terms diagonal arrangement reflection arrangement The transducers are mounted on opposite sides of the pipe. The transducers are mounted on the same side of the pipe.
  • Page 16 4 Basics – flow measurement 4.2 Measurement arrangements PIOX S72* Transducer distance The transducer distance is measured between the inner edges of the transducers. reflection arrangement diagonal arrangement (positive transducer distance) diagonal arrangement (negative transducer distance) a – transducer distance Sound beam plane Plane containing 1 or several sound paths or beams Fig.
  • Page 17: Acoustic Penetration

    4 Basics – flow measurement PIOX S72* 4.3 Acoustic penetration 4.2.2 Examples diagonal arrangement with 1 beam reflection arrangement with 1 beam 1 transducer pair 1 transducer pair 1 sound path 2 sound paths 1 beam 1 beam 1 plane 1 plane diagonal arrangement with 2 beams reflection arrangement with 2 beams and 2 planes...
  • Page 18: Undisturbed Flow Profile

    4 Basics – flow measurement 4.4 Undisturbed flow profile PIOX S72* Observe the following notes on the selection of the measuring point: Horizontal pipe Select a measuring point where the transducers can be mounted laterally on the pipe, allowing the sound waves to propagate horizontally in the pipe.
  • Page 19 4 Basics – flow measurement PIOX S72* 4.4 Undisturbed flow profile Tab. 4.1: Recommended distance from disturbances D – nominal pipe diameter at the measuring point l – recommended distance between disturbance and transducer position disturbance: 90° elbow inlet: l ≥ 10 D outlet: l ≥...
  • Page 20 4 Basics – flow measurement 4.4 Undisturbed flow profile PIOX S72* Tab. 4.1: Recommended distance from disturbances D – nominal pipe diameter at the measuring point l – recommended distance between disturbance and transducer position disturbance: valve inlet: l ≥ 40 D outlet: l ≥...
  • Page 21: Product Description

    5 Product description PIOX S72* 5.1 Measuring system Product description Measuring system The measuring system consists of the transmitter, the temperature probe, the ultrasonic transducers and the pipe on which the measurement is taken. Fig. 5.1: Example of a measurement arrangement 1 –...
  • Page 22 5 Product description 5.2 Handling concept PIOX S72* Tab. 5.1: Description of the program branches program branch description Before starting a measurement, the transducer, pipe and fluid parameters have to be entered in the Parameters program branch Parameters. After activating the measuring channels and the input of the transducer distance, the measurement is Measurement started in the program branch Measurement.
  • Page 23: Display

    5 Product description PIOX S72* 5.3 Display Display Structure Fig. 5.3: Menu item of the program branch Parameters 1 – program branch 2 – menu item currently edited 3 – area for scroll lists, selection fields or input fields Tab. 5.2: Navigation horizontal scroll list vertical scroll list...
  • Page 24: Keyboard

    5 Product description 5.4 Keyboard PIOX S72* Keyboard The keyboard has 15 keys, including 3 function keys: ENTER, and C. Some keys have multiple functions. They can be used to enter data, to navigate through scroll lists as well as to execute special functions (e.g., reset of totalizers).
  • Page 25: Transport And Storage

    The measuring equipment must be packaged properly for transport. For weight indications, see technical specification. • Use, if possible, the original packaging by FLEXIM or an equivalent cardboard box. • Position the transmitter, transducers and accessories in the middle of the cardboard box.
  • Page 26: Mounting

    7 Mounting PIOX S72* Mounting Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 27: Transmitter

    7 Mounting PIOX S72* 7.1 Transmitter Important! Remove all protective foils before using the measuring equipment in explosive atmospheres (see document SIFLUXUS). Transmitter 7.1.1 Opening and closing the housing 7.1.1.1 Opening Caution! Possible danger by opening the equipment at improper ambient conditions Open the transmitter in safe ambient conditions only (e.g., air humidity <...
  • Page 28 7 Mounting 7.1 Transmitter PIOX S72* Transmitter with aluminum housing • Close the housing cover pressing it slightly upwards on the side of the screws. • Hand-tight all screws of the transmitter housing. Fig. 7.2: Transmitter 1 – screws 7.1.2 Installation of the transmitter Notice! Install the equipment in a shady place, the cable glands facing downward.
  • Page 29 7 Mounting PIOX S72* 7.1 Transmitter Transmitter with aluminum housing • Fix the wall mount (2) with 3 countersunk-head screws (3) firmly to the wall. • Attach the transmitter (1) to the lower hooks (7) of the wall mount. • Fix the transmitter to the wall mount by locking the stop bolt (4) into the destined hole and by firmly tightening the screw (5).
  • Page 30 7 Mounting 7.1 Transmitter PIOX S72* 7.1.2.2 Pipe mounting Notice! The pipe has to be sufficiently stable to withstand the pressure exerted by the transmitter and the shackles. Transmitter with stainless steel housing Installation on a 2'' pipe • Position the transmitter and the shackle (1) on the pipe. •...
  • Page 31 7 Mounting PIOX S72* 7.1 Transmitter Transmitter with aluminum housing Notice! For the mounting on pipes the pipe mounting kit, consisting of 2 band clamps, 2 countersunk-head screws and 2 washers, is needed. The distance of the band clamps has to be identical to the distance of the holes on the wall mount.
  • Page 32: Transducers

    7 Mounting 7.2 Transducers PIOX S72* Transducers Mounting Caution! Warning of severe injuries from hot or cold components Touching hot or cold components can lead to severe injuries (e.g., thermal damage). → Any mounting, installation or connection work has to be concluded. →...
  • Page 33 7 Mounting PIOX S72* 7.2 Transducers Rust, paint or deposits on the pipe absorb the sound signal. A good acoustic contact between the pipe and the transducers is obtained as follows: • Clean the pipe at the selected measuring point. –...
  • Page 34 7 Mounting 7.2 Transducers PIOX S72* Diagonal arrangement with 2 beams Reflection arrangement with 2 beams and 2 planes X arrangement • same characteristics as reflection arrangement with 1 beam • additional characteristic: influences of the flow profile are compensated because the measurement takes place in 2 planes displaced X arrangement •...
  • Page 35 7 Mounting PIOX S72* 7.2 Transducers Bidirectional measurement Measurement in reflection arrangement with 2 beams and 2 planes flow direction flow direction • The sound beam plane is selected • The 2 sound beam planes are selected according to the nearest elbow at an angle of 45°...
  • Page 36 7 Mounting 7.2 Transducers PIOX S72* 7.2.2.2 Transducer arrangement The transducers can be arranged in the mounting rails in different ways: Fig. 7.9: Transducer arrangement in mounting rails reflection arrangement, 1 rail reflection arrangement, 2 rails diagonal arrangement, 2 parallel rails diagonal arrangement, 2 displaced rails 7.2.2.3...
  • Page 37 7 Mounting PIOX S72* 7.2 Transducers 7.2.2.4 Mounting with Variofix L (PermaRail) Scope of delivery (example) Variofix L transducer pair quick release clasp with tension strap band clamp clasp with tension strap ratchet clasp and tension strap coil UMPIOX_S72xV1-9EN, 2022-05-15...
  • Page 38 7 Mounting 7.2 Transducers PIOX S72* Mounting When measuring in diagonal arrangement, the transducer mounting fixtures are mounted on opposite sides of the pipe. When measuring in reflection arrangement, the transducer mounting fixtures are mounted on the same side of the pipe. When measuring in diagonal arrangement with 2 beams in displaced X arrangement, 4 transducer mounting fixtures have to be mounted.
  • Page 39 7 Mounting PIOX S72* 7.2 Transducers Step 1: Disassembly of the transducer mounting fixture Variofix L • Disassemble the transducer mounting fixture Variofix L. Fig. 7.13: Disassembly of the transducer mounting fixture Variofix L 1 – cover 2 – screw 3 –...
  • Page 40 7 Mounting 7.2 Transducers PIOX S72* • Insert approx. 100 mm of the tension strap into part (1) and (2) of the ratchet clasp (see Fig. 7.16 a). • Bend the tension strap. • Insert the tension strap into part (1) of the ratchet clasp (see Fig. 7.16 b). •...
  • Page 41 7 Mounting PIOX S72* 7.2 Transducers Fig. 7.18: Tension strap with clamp Fig. 7.19: Tension strap with quick release clasp and metal spring 1 – tension strap clamp 1 – clasp screw 2 – metal spring Fig. 7.20: Tension strap with band clamp clasp 1 –...
  • Page 42 7 Mounting 7.2 Transducers PIOX S72* Fig. 7.21: Tension strap with metal spring and clamp 1 – metal spring 2 – tension strap clamp Fig. 7.22: Ratchet clasp with tension strap 1 – sense of rotation 2 – edge 3 – lever 4 –...
  • Page 43 7 Mounting PIOX S72* 7.2 Transducers • Select the installation instruction of the supplied clasp: Band clamp clasp • Insert the tension strap into the tension strap clamp (2). • Place the tension strap around the pipe and insert it into the clasp (see Fig. 7.25 and Fig. 7.26). •...
  • Page 44 7 Mounting 7.2 Transducers PIOX S72* Ratchet clasp • Insert the tension strap into its clamp (2) and the metal spring (see Fig. 7.25 and Fig. 7.28). The metal spring does not have to be mounted on: – steel pipes –...
  • Page 45 7 Mounting PIOX S72* 7.2 Transducers • Repeat the steps to fix the second rail, see Fig. 7.30. Fig. 7.30: Pipe with 2 rails Step 5: Installation of the transducers in the mounting fixture Variofix L • Press the transducers firmly into their clamping fixtures in the covers until they are tightly fixed. The transducer cables show in opposite directions (see Fig.
  • Page 46 7 Mounting 7.2 Transducers PIOX S72* Fig. 7.32: Adjustment of the transducer distance 1 – cover 2 – strain relief clamp 3 – equipotential bonding terminal a – transducer distance Notice! Make sure that the coupling foil remains on the contact surface of the transducers. For information concerning the coupling foil, see the safety data sheet.
  • Page 47 7 Mounting PIOX S72* 7.2 Transducers 7.2.2.5 Mounting with Variofix C Scope of delivery (example) Variofix C transducer pair quick release clasp with tension strap band clamp clasp with tension strap tension strap coil ratchet clasp UMPIOX_S72xV1-9EN, 2022-05-15...
  • Page 48 7 Mounting 7.2 Transducers PIOX S72* Mounting When measuring in reflection arrangement, 1 transducer mounting fixture is mounted laterally on the pipe. When measuring in diagonal arrangement, 2 transducer mounting fixtures are mounted on opposite sides of the pipe. In the following, the installation of 1 transducer mounting fixture in reflection arrangement is described. Fig.
  • Page 49 7 Mounting PIOX S72* 7.2 Transducers Fig. 7.37: Disassembly of the transducer mounting fixture Variofix C 1 – cover 2 – tensioning screw 3 – spring clip 4 – indentation 5 – spacing element 6 – rail 7 – tension strap clamp Step 2: Fixation of the clasps to the tension straps •...
  • Page 50 7 Mounting 7.2 Transducers PIOX S72* Ratchet clasp • Cut the tension strap to length (pipe circumference + at least 120 mm). Caution! The edge of the tension strap is very sharp. Risk of injury. → Debur sharp edges. → Wear the required personal protective equipment. →...
  • Page 51 7 Mounting PIOX S72* 7.2 Transducers Quick release clasp • Insert the tension strap into its clamp and the metal spring (see Fig. 7.42 and Fig. 7.43). • Position the clasp, the tension strap clamp and the metal spring on the pipe (see Fig. 7.41): –...
  • Page 52 7 Mounting 7.2 Transducers PIOX S72* Notice! In order to release the screw and the tension strap, press the lever down (see Fig. 7.46). Fig. 7.45: Tension strap with metal spring and clamp 1 – metal spring 2 – tension strap clamp Fig.
  • Page 53 7 Mounting PIOX S72* 7.2 Transducers Fig. 7.48: Rail, fixed on one side to the pipe 1 – tension strap clamp 2 – tension strap clamp 3 – nut • Select the installation instruction of the supplied clasp: Band clamp clasp •...
  • Page 54 7 Mounting 7.2 Transducers PIOX S72* Ratchet clasp • Insert the tension strap into its clamp (2) and the metal spring (see Fig. 7.49 and Fig. 7.52). The metal spring does not have to be mounted on: – steel pipes –...
  • Page 55 7 Mounting PIOX S72* 7.2 Transducers Step 5: Installation of the transducers in the mounting fixture Variofix C • Put coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the transducers. The coupling foil can be fixed to the contact surface with a small amount of coupling compound. Notice! If the signal is not sufficient for the measurement, use coupling compound instead of coupling foil.
  • Page 56 7 Mounting 7.2 Transducers PIOX S72* Fig. 7.56: Variofix C with transducers on the pipe 1 – cover 2 – screw Remove the cover from the mounted transducer mounting fixture Variofix C as follows: • Loosen the screws (see Fig. 7.56). •...
  • Page 57 7 Mounting PIOX S72* 7.2 Transducers 7.2.2.6 Mounting with PermaLok Scope of delivery 2 × PermaLok 1 × transducer pair 4 × quick release clasp with tension strap Fig. 7.58: PermaLok components 1 – cover 2 – screw 3 – nut 4 –...
  • Page 58 7 Mounting 7.2 Transducers PIOX S72* Mounting When measuring in diagonal arrangement, the transducer mounting fixtures are mounted on opposite sides of the pipe, see Fig. 7.59. When measuring in reflection arrangement, the transducer mounting fixtures are mounted on the same side of the pipe, see Fig.
  • Page 59 7 Mounting PIOX S72* 7.2 Transducers Step 2: Fixation of the rail to the pipe • Insert the tension strap through the slot of the tension strap clamp (if present, see Fig. 7.62) or through the rail fixture, see Fig. 7.63. Fig.
  • Page 60 7 Mounting 7.2 Transducers PIOX S72* Step 3: Mounting of the transducers in the mounting fixture PermaLok • Stick coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the transducer. The coupling foil can be fixed to the contact surface with a small amount of coupling compound. •...
  • Page 61 7 Mounting PIOX S72* 7.2 Transducers 7.2.2.7 Mounting with tension strap and quick release clasp Caution! The edge of the tension strap is very sharp. Risk of injury. → Debur sharp edges. → Wear the required personal protective equipment. → Observe the applicable rules. •...
  • Page 62 7 Mounting 7.2 Transducers PIOX S72* 7.2.2.8 Fixation of PermaLok with solid band strap kit • Cut the tension strap to length (pipe circumference + at least 305 mm). • Insert the solid band through the slot of the tension strap clamp. •...
  • Page 63 7 Mounting PIOX S72* 7.2 Transducers • Hold the tabs of the clip using pliers and pull the solid band completely through the slot of the clip, see Fig. 7.73. • Bend the end of the solid band back over the set of teeth, see Fig. 7.74. Fig.
  • Page 64: Temperature Probe

    7 Mounting 7.3 Temperature probe PIOX S72* Temperature probe Mounting 7.3.1 Pipe preparation Caution! Contact with grinding dust This may result in injuries (e.g., breathing difficulties, skin reactions, eye irritations). → Wear the required personal protective equipment. → Observe the applicable rules. Important! The pipe has to be sufficiently stable to withstand the pressure exerted by the temperature probe.
  • Page 65 7.3 Temperature probe Fig. 7.77: Clasp Fig. 7.78: Temperature probe on the pipe 7.3.2.2 Installation with FLEXIM clasp Caution! The edge of the tension strap is very sharp. Risk of injury! → Debur sharp edges. → Wear the required personal protective equipment.
  • Page 66 7 Mounting 7.3 Temperature probe PIOX S72* 7.3.2.3 Installation with quick release clasp Caution! The edge of the tension strap is very sharp. Risk of injury! → Debur sharp edges. → Wear the required personal protective equipment. → Observe the applicable rules. •...
  • Page 67: Connection

    8 Connection PIOX S72* Connection Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 68: Transducers

    8 Connection 8.1 Transducers PIOX S72* Transducers It is recommended to run the cables from the measuring point to the transmitter before connecting the transducers to avoid load on the connectors. Notice! If transducers are replaced or added, the SENSPROM has to be replaced or added as well. Fig.
  • Page 69 8 Connection PIOX S72* 8.1 Transducers Fig. 8.2: Connection of the transducer cable with SMB connectors to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – SMB connector 2 – cable gland 8.1.1.2 Transducer cable with plastic jacket and stripped ends •...
  • Page 70 8 Connection 8.1 Transducers PIOX S72* Fig. 8.3: Connection of the transducer cable with plastic jacket and stripped ends to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE cable gland X_AV X_AR X_BV X_BR 1 – cap nut 2 –...
  • Page 71 8 Connection PIOX S72* 8.1 Transducers Fig. 8.4: Connection of the transducer cable with stainless steel conduit and stripped ends to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – cable gland 8.1.2 Connection of the extension cable to the transmitter The extension cable is connected to the transmitter via the transducer connection.
  • Page 72 8 Connection 8.1 Transducers PIOX S72* Fig. 8.5: Connection of the extension cable to the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE cable gland X_AV X_AR X_BV X_BR 1 – cap nut 2 – compression part 3 – basic part 4 –...
  • Page 73 8 Connection PIOX S72* 8.1 Transducers Fig. 8.6: Connection of the transducer cable with SMB connectors 1 – SMB connector 2 – cable gland 8.1.3.2 Transducer cable with plastic jacket and stripped ends • Remove the blind plug for the connection of the transducer cable. •...
  • Page 74 8 Connection 8.1 Transducers PIOX S72* Fig. 8.7: Connection of the transducer cable with plastic jacket and stripped ends cable gland 1 – cap nut 2 – compression part 3 – basic part 4 – external shield, brushed back 8.1.3.3 Transducer cable with stainless steel conduit and stripped ends •...
  • Page 75 8 Connection PIOX S72* 8.1 Transducers Fig. 8.8: Connection of the transducer cable with stainless steel conduit and stripped ends 1 – cable gland 8.1.4 Connection of the extension cable to the junction box 8.1.4.1 Connection without potential separation (standard) The connection of the extension cable to the junction box without potential separation ensures that the transducer, junction box and transmitter are on the same potential.
  • Page 76 8 Connection 8.1 Transducers PIOX S72* Tab. 8.8: Terminal assignment terminal connection (extension cable) white or marked cable (core) white or marked cable (internal shield) brown cable (internal shield) brown cable (core) cable gland external shield Fig. 8.9: Connection of the extension and transducer cable to the junction box cable gland 1 –...
  • Page 77 8 Connection PIOX S72* 8.1 Transducers 8.1.4.2 Connection with potential separation If earthing on the same potential cannot be ensured, e.g., in measurement arrangements with long extension cables, the extension cable and the junction box have to be electrically insulated from each other. The junction box and the transducers have to be on the same potential.
  • Page 78 8 Connection 8.1 Transducers PIOX S72* Fig. 8.10: Connection of the extension and transducer cable to the junction box cable gland 1 – extension cable 2 – external shield 3 – shield terminal 4 – cap nut 5 – compression part 6 –...
  • Page 79 8 Connection PIOX S72* 8.1 Transducers 8.1.5 SENSPROM The SENSPROM contains important transducer data for the operation of the transmitter with the transducers. If transducers are replaced or added, the SENSPROM has to be replaced or added as well. Notice! The serial numbers of the SENSPROM and the transducer have to be identical.
  • Page 80: Power Supply

    8 Connection 8.2 Power supply PIOX S72* Power supply Connection The installation of the power supply is carried out by the operator. The operator has to provide an overcurrent protector (fuse or similar device) disconnecting all energizing wires in case of an inadmissible high current consumption. The impedance of the protective earth has to be low ohmic in order not to allow touch voltage pass the permissible limit.
  • Page 81 8 Connection PIOX S72* 8.2 Power supply Tab. 8.10: Terminal assignment terminal connection AC connection DC earth earth N(-) neutral L(+) phase 100...230 V AC, 50...60 Hz fuse 1 A, time-lag 1.6 A, time-lag 8.2.1 Cable connection Transmitter with stainless steel housing •...
  • Page 82: Outputs

    8 Connection 8.3 Outputs PIOX S72* Fig. 8.15: Cable gland for aluminum housing 1 – cap nut 2 – compression part 3 – basic part Outputs Important! The max. permissible voltage between the outputs and against PE is 60 V DC (permanent). •...
  • Page 83 8 Connection PIOX S72* 8.3 Outputs Tab. 8.11: Output circuits output transmitter external circuit explanation internal circuit connection passive current output/HART current output = 8…30 V > 0.024 A [Ω] + 8 V example: = 30 V ≤ 900 Ω HART = 10…24 V switchable current output...
  • Page 84 8 Connection 8.3 Outputs PIOX S72* Tab. 8.11: Output circuits output transmitter external circuit explanation internal circuit connection binary output (optorelay) circuit 1 ≤ 26 V TF7- ≤ 100 mA *721**-******-******a-*****-** [kΩ] = U [mA] a = U, V, D circuit 2 binary output (Reed relay) circuit 1...
  • Page 85: Inputs

    8 Connection PIOX S72* 8.4 Inputs Tab. 8.11: Output circuits output transmitter external circuit explanation internal circuit connection digital output circuit 1 = 5…30 V TF7- < 100 mA *72***-******-******a-*****-** [kΩ] = U [mA] a = M, P, Q circuit 2 RS485 120 Ω...
  • Page 86 8 Connection 8.4 Inputs PIOX S72* 8.4.1 Current input An active (self-powered) or a passive (external powered) current source can be connected to the current inputs of the transmitter. Connection of an active current source Important! The terminals Txa and Txb must not be connected. For the connection of the input cable to the transmitter see section 8.2.1, Fig.
  • Page 87 8 Connection PIOX S72* 8.4 Inputs 8.4.2 Binary input The transmitter can be equipped with max. 4 binary inputs. Binary inputs can process measured values with short pulse. Via the binary inputs, it is possible to remotely trigger some functions of the transmitter. For the connection of the input cable to the transmitter, see section 8.2.1, Fig.
  • Page 88: Temperature Probe

    8 Connection 8.5 Temperature probe PIOX S72* Temperature probe Connection It is possible to connect the temperature probes Pt100/Pt1000 (4-wire) to the inputs of the transmitter (optional). Fig. 8.18: Connection of the temperature probes to the transmitter NEXT DISP SNAP DISP 3x Q FAST...
  • Page 89 8 Connection PIOX S72* 8.5 Temperature probe 8.5.1 Direct connection of the temperature probe • Remove the blind plug for the connection of the temperature probe. • Open the cable gland of the temperature probe. The compression part remains in the cap nut. •...
  • Page 90 8 Connection 8.5 Temperature probe PIOX S72* 8.5.2 Connection with extension cable • Remove the blind plug for the connection of the temperature probe. • Open the cable gland of the extension cable. The compression part remains in the cap nut. •...
  • Page 91 8 Connection PIOX S72* 8.5 Temperature probe Fig. 8.21: Preparation of the extension cable 110 mm 7 mm 1 – cap nut 2 – compression part 3 – basic part 4 – reducer 5 – external shield, brushed back 6 – sealing ring side Fig.
  • Page 92: Service Interfaces

    8 Connection 8.6 Service interfaces PIOX S72* Service interfaces Connection 8.6.1 USB interface The transmitter can be connected directly to the PC via the USB interface. • Connect the USB cable to the USB interface of the transmitter and to the PC. Fig.
  • Page 93 8 Connection PIOX S72* 8.6 Service interfaces Fig. 8.25: Cable gland 1 – cap nut 2 – compression part 3 – basic part 4 – sealing ring (only for cable gland M20, not for cable gland 1/2 NPS) 5 – housing wall 6 –...
  • Page 94 8 Connection 8.6 Service interfaces PIOX S72* Fig. 8.27: Cable gland 1 – cap nut 2 – compression part 3 – basic part • Push the cable through the cap nut, compression part and basic part of the cable gland. •...
  • Page 95: Start-Up

    9 Start-up PIOX S72* Start-up Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 96: Start-Up Settings

    9 Start-up 9.1 Start-up settings PIOX S72* Start-up settings When starting up the transmitter for the first time, the following settings are required: • language • time/date • unit of measurement These displays will only be indicated when the transmitter is switched on for the first time or after an initialization. Language The available transmitter languages are displayed.
  • Page 97: Status Indications

    9 Start-up PIOX S72* 9.3 Status indications Status indications The operation state is indicated by LEDs. Fig. 9.1: Command panel of the transmitter NEXT DISP SNAP DISP 3x Q FAST MODE X_AV X_AR X_BV X_BR 1 – LED (channel A) 2 –...
  • Page 98: Program Branches

    9 Start-up 9.4 Program branches PIOX S72* Program branches The following schema shows the program branches. For a detailed overview of the menu structure see annex A. Parameters Measurement Options Special functions Selection of a measuring channel Selection of measuring channels Selection of a measuring channel System settings Transducer parameters...
  • Page 99: Date And Time

    9 Start-up PIOX S72* 9.7 Date and time It will be requested whether the initial settings are to be carried out. Initial settings If Yes is selected, the following setting dialogs will be displayed. •Language •Date/time •Units of measurement •Delete meas. values •Delete snaps •...
  • Page 100: Measurement

    10 Measurement 10.1 Enabling the analysis PIOX S72* Measurement Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 101: Parameter Input

    10 Measurement PIOX S72* 10.2 Parameter input Special functions\Measurement\Measurement modes\Reference pressure • Enter the pressure for the local reference conditions (default value: 1 bar). • Press ENTER. 10.2 Parameter input Notice! Avoid a simultaneous parameter entry via the keyboard and the USB, LAN or process interface. The parameter records received via these interfaces will overwrite the current transmitter parameterization.
  • Page 102 10 Measurement 10.2 Parameter input PIOX S72* Parameters\Connected transd. The parameters can also be entered without connected transducers or SENSPROM. • Select the list item Connected transd.. • Press ENTER. The display Transducer not found will be indicated if neither transducers nor SENSPROM are connected. •...
  • Page 103 10 Measurement PIOX S72* 10.2 Parameter input Sound speed of the pipe material Parameters\Pipe material\Other material\c material • Enter the sound speed of the pipe material. Notice! There are 2 sound speeds for pipe materials: the longitudinal and the transversal one. Enter the sound speed which is nearer to 2500 m/s.
  • Page 104 10 Measurement 10.2 Parameter input PIOX S72* Sound speed of the lining material Parameters\Lining material\Other material\c material • Enter the sound speed of the lining material. Notice! For pipe materials there are 2 sound speeds, the longitudinal and the transversal one. Enter the sound speed which is nearer to 2500 m/s.
  • Page 105 10 Measurement PIOX S72* 10.2 Parameter input 10.2.3 Input of fluid parameters Fluid Notice! If analysis is enabled as measuring mode, the scroll list will only display fluid data sets which can output a physical quantity for analysis. Parameters\Fluid • Select the fluid from the scroll list. •...
  • Page 106 10 Measurement 10.2 Parameter input PIOX S72* Fluid temperature Parameters\Fluid temp. The fluid temperature is used: – at the beginning of the measurement for the interpolation of the sound speed and therefore for the calculation of the recommended transducer distance –...
  • Page 107: Measurement Settings

    10 Measurement PIOX S72* 10.3 Measurement settings 10.2.5 Other parameters Extension cable Parameters\Extension cable In case the transducer cable is extended (e.g., between junction box and transmitter), enter the length of the extension cable. • Select the menu item Extension cable in the program branch Parameters. •...
  • Page 108 10 Measurement 10.3 Measurement settings PIOX S72* 10.3.3 Input of the damping factor Options\...\Damping Each displayed measured value is a floating average of the last x seconds, with x being the damping factor. If 0 s is entered as damping factor, no average is calculated. The value of 10 s is appropriate for normal flow conditions.
  • Page 109 10 Measurement PIOX S72* 10.3 Measurement settings 10.3.6 Configuration of an output If the transmitter is equipped with outputs, they have to be configured. The measured value, the status value or an Options event value can be transmitted via the different outputs. Channel x In the following, the configuration of an analog output is described.
  • Page 110 10 Measurement 10.3 Measurement settings PIOX S72* Tab. 10.1: Configuration of the outputs source item list item output physical quantity for analysis selected in the program branch Parameters Flow/analysis Mass fraction (here: Mass fraction) flow velocity Flow velocity volumetric flow rate Volumetric flow rate mass flow rate...
  • Page 111 10 Measurement PIOX S72* 10.3 Measurement settings Tab. 10.1: Configuration of the outputs source item list item output measured values of input quantities (e.g., temperature, density) which are Miscellaneous Custom. Input 1 not used for calculation Custom. Input 2 In the menu item Options\Assign inputs it is possible to assign configured inputs to customized inputs.
  • Page 112 10 Measurement 10.3 Measurement settings PIOX S72* Error output Options\Channel A\Outputs\...\Error value An error value can be defined which is output if the source item cannot be measured. • Select a list item for the error output. • Press ENTER. •...
  • Page 113 10 Measurement PIOX S72* 10.3 Measurement settings Tab. 10.3: Examples for the error output (output range: 4...20 mA) list item output signal 20.0 mA I [mA] Other value I [mA] error value = 3.5 mA Measuring range The sign of the measured value and the measuring range are determined. Options\Channel A\Outputs\...\Measured values\Absolute value •...
  • Page 114 10 Measurement 10.3 Measurement settings PIOX S72* Output function test The function of the output can now be tested. • Connect an external measuring instrument to the terminals of the output. Options\Channel A\Outputs\...\Test signal • Select Yes to test the output. Select No to display the next menu item. •...
  • Page 115 10 Measurement PIOX S72* 10.3 Measurement settings status value – status OK event value – idle state The status of the output signal is defined which is to be output The status of the output signal is defined which is to be output if when measuring a measured value.
  • Page 116: Start Of The Measurement

    10 Measurement 10.4 Start of the measurement PIOX S72* 10.3.6.3 Switchable current outputs If the transmitter has switchable current outputs, it has to be defined how to switch them. Special functions • Select the menu item Current output in the program branch Special functions.
  • Page 117 10 Measurement PIOX S72* 10.4 Start of the measurement Input of the measuring point number Measurement\Measuring point no. • Enter the number of the measuring point. • Press ENTER. For the activation of text input see Special functions\Dialogs/Menus\Measuring point no. Input of the sound path number Measurement\Sound path •...
  • Page 118 10 Measurement 10.4 Start of the measurement PIOX S72* Tab. 10.4: Diagnostic values display explanation line 1 of the display SCNR signal-to-correlated noise ratio scroll with signal-to-noise ratio signal quality By pressing it is possible to display the numeric value instead of the bar graph.
  • Page 119: Field Calibration

    10 Measurement PIOX S72* 10.5 Field calibration Tab. 10.6: Max. permissible deviation between the recommended and entered transducer distance transducer frequency (3rd max. difference between the recommended and entered transducer distance [mm] character of the technical type) shear wave transducer Lamb wave transducer -60…+120 -45…+90...
  • Page 120 10 Measurement 10.5 Field calibration PIOX S72* Stop measurement • Press during the measurement. The following display appears: Fig. 10.4: Scroll list in the program branch Measurement • Select Stop measurement. • Press ENTER. The measurement is stopped. Input of the reference value •...
  • Page 121: Display Of Measured Values

    10 Measurement PIOX S72* 10.6 Display of measured values Auto calibration • Select the list item Auto calibration. • Press ENTER. The calibration is carried out. Notice! If several reference values were stored, e.g., when realizing diverse samplings with different concentrations, an average is made during the auto calibration which is valid for the entire operating range.
  • Page 122 10 Measurement 10.6 Display of measured values PIOX S72* Toggling between the channels If the measurement is started on several channels, the display of the measured values can be adapted as follows: AutoMux mode If the AutoMux mode is activated, the measured values of all activated channels (measuring and calculation channels) are displayed consecutively.
  • Page 123 10 Measurement PIOX S72* 10.6 Display of measured values Tab. 10.7: Description of the status line value explanation flow velocity comparison of the measured flow velocity with the flow limits of the system √ OK, the flow velocity is not within the critical range ↑...
  • Page 124: Display Of Parameters

    10 Measurement 10.7 Display of parameters PIOX S72* 10.7 Display of parameters The parameters can be displayed during the measurement. • Press during the measurement. The following display appears: Fig. 10.10: Scroll list in the program branch Measurement 1 – status indication The measurement is running in the background.
  • Page 125: Repeated Display Of Measured Values

    10 Measurement PIOX S72* 10.8 Repeated display of measured values 10.8 Repeated display of measured values • Select the program branch Measurement to return to the measured value display. • Press ENTER. The following display appears: Fig. 10.12: Scroll list in the program branch Measurement Measurement\Show measurement •...
  • Page 126 10 Measurement 10.9 Execution of special functions PIOX S72* display description reset the totalizer to zero Reset totalizers display the measured value of the totalizer for several seconds Freeze display reset the totalizer error Fehler zurücksetzen stop totalizer and reset it to zero Stop/clear totalizers start totalizer Start totalizers...
  • Page 127: Stop Of The Measurement

    10 Measurement PIOX S72* 10.10 Stop of the measurement 10.10 Stop of the measurement • Press during the measurement. The following display appears: Fig. 10.13: Scroll list in the program branch Measurement Measurement\Stop measurement • Select Stop measurement. • Press ENTER. The measurement is stopped.
  • Page 128: Troubleshooting

    11 Troubleshooting PIOX S72* Troubleshooting Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 129: Problems With The Measurement

    The signal is lost during the measurement. • If there is no measuring signal after the pipe had been run empty and refilled, contact FLEXIM. • Wait a moment until the acoustic contact is reestablished. The measurement can be interrupted due to a temporarily higher proportion of gas bubbles and solids in the fluid.
  • Page 130 • field calibration according to section 10.5 c>SPEEDBOUND limit • adaptation of the internal application data base by FLEXIM sound speed is less than the lower limit • field calibration according to section 10.5 c>SPEEDBOUND •...
  • Page 131: Measuring Point Selection

    Measurements on pipes made of porous materials (e.g., concrete or cast iron) are only conditionally possible • Contact FLEXIM. The pipe lining may cause problems during the measurement if it is not firmly attached to the inner pipe wall or consists of an acoustically absorbing material •...
  • Page 132: Significant Deviations Of The Measured Values

    11 Troubleshooting 11.5 Significant deviations of the measured values PIOX S72* 11.5 Significant deviations of the measured values A fluid with a wrong sound speed was selected. • If a fluid was selected whose sound speed does not match the actual one, a pipe wall signal can be mistaken for the measuring signal.
  • Page 133: Maintenance And Cleaning

    12 Maintenance and cleaning PIOX S72* Maintenance and cleaning Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 134: Maintenance

    • the transducers were used for a prolonged period at high temperatures (several months > 130 °C for normal transducers or > 200 °C for high temperature transducers) In order to realize a recalibration under reference conditions, either the transmitter, the transducers or both have to be sent to FLEXIM. 12.4 Firmware update A firmware update is carried out with FluxDiagReader or FluxDiag (option).
  • Page 135: Dismounting And Disposal

    13 Dismounting and disposal PIOX S72* Dismounting and disposal Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (ATEX, IECEx) This may result in personal or material damage or dangerous situations. → Observe the "Safety instructions for the use in explosive atmospheres" (see document SIFLUXUS). Danger! Risk of explosion when using the measuring equipment in explosive atmospheres (TR TS) This may result in personal or material damage or dangerous situations.
  • Page 136: Dismounting

    Batteries must be disposed of separately from electrical or electronic equipment. For this purpose, remove the batteries from the device and take them to the designated disposal system. The components are taken back free of charge by FLEXIM in accordance with national regulations. Contact FLEXIM. 2022-05-15, UMPIOX_S72xV1-9EN...
  • Page 137: User Modes

    14 User modes PIOX S72* User modes The user modes allow extended diagnostics of signals and measured values as well as the definition of additional Special functions parameters adapted to the application. The following user modes can be selected: System settings •StandardUser •ExpertUser User mode...
  • Page 138: Standarduser Mode

    14 User modes 14.1 StandardUser mode PIOX S72* User mode selection Special functions\System settings\User mode • Select the menu item User mode. • Press ENTER. • Select a list item. • Press ENTER. Special settings Options\Channel\Special settings • Select the measuring channel in the program branch Options. •...
  • Page 139 14 User modes PIOX S72* 14.1 StandardUser mode 14.1.2 Turbulence mode In the presence of high turbulence, i.e., high Reynolds numbers or disturbed flow profiles due to short inlet and outlet lengths, there are large fluctuations in the transit time of the ultrasonic signals, resulting in poor signal quality (e.g., reduction in signal amplitude, increase in gain).
  • Page 140: Expertuser Mode

    14 User modes 14.2 ExpertUser mode PIOX S72* Options\Special settings\Multi-point calibration\Bidirectional use • Select Yes to apply the calibration curve for negative flow velocities as well. Select No if it is not to be used for negative flow velocities. 14.2 ExpertUser mode Some menu items that are not visible in the StandardUser mode are displayed.
  • Page 141 14 User modes PIOX S72* 14.2 ExpertUser mode Fig. 14.1: Flow velocity outside the valid range 1 – physical quantity 2 – flow velocity Notice! If the limit of the flow velocity +Flow velocity limit is too low or -Flow velocity limit is too high, a measurement might be impossible because most of the measured values will be marked as invalid.
  • Page 142 14 User modes 14.2 ExpertUser mode PIOX S72* Options\Special settings\Relative threshold • Enter the value for the relative threshold for the particular measuring channel. • Press ENTER. Example absolute threshold: 2007 m/s relative threshold: 600 m/s value of the sound speed curve at the measuring point: 1546 m/s As 1546 m/s + 600 m/s = 2146 m/s is greater than the absolute value 2007 m/s, this value will be used as the absolute limit of the sound speed when the plausibility of the signal is evaluated.
  • Page 143: Superuser Mode And Superuser Ext. Mode

    14 User modes PIOX S72* 14.3 SuperUser mode and SuperUser ext. mode Tab. 14.2: Disturbance parameters disturbance type further inputs description Disturbance distance (l) 90° elbow Disturbance distance (l Double elbow Dist. between elbows (l Disturbance distance (l Double elb. out of plane Dist.
  • Page 144 14 User modes 14.3 SuperUser mode and SuperUser ext. mode PIOX S72* Notice! Some of the defined parameters remain activated when switching to the StandardUser mode. These parameters are displayed but cannot be changed. Notice! If the analysis is activated, not all settings are applicable. 14.3.1 Pig detection This function detects pigs inside the pipe.
  • Page 145 14 User modes PIOX S72* 14.3 SuperUser mode and SuperUser ext. mode 14.3.3 Linear calibration It is possible to define a correction of the flow velocity: = m · v + n where – measured flow velocity – factor, range: -2...+2 –...
  • Page 146 14 User modes 14.3 SuperUser mode and SuperUser ext. mode PIOX S72* 14.3.4 Weighting factor The weighting factor is used for transducers installed on the same pipe in order to compensate differences between the measured values of the flow velocity of different channels. The differences can be caused by profile deformations or cross-flows.
  • Page 147: Outputs (Tf7-*721**-******-******A-*****-**; A = U, V, D)

    15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) PIOX S72* 15.1 Configuration of a binary output Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) If the transmitter is equipped with outputs, they have to be configured. For the configuration of the analog output, see Options section 10.3.6.
  • Page 148 15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) 15.1 Configuration of a binary output PIOX S72* Assignment of a source item Options\Channel A\Outputs\...\Source item Depending on the selected source item, status or event values can be output. Tab. 15.1: Output of status values or event values source item status value event value...
  • Page 149 15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) PIOX S72* 15.1 Configuration of a binary output Output function test The function of the output can now be tested. • Connect a multimeter to the output. Options\Channel A\Outputs\...\B1 Test signal • Select Yes to test the output. Select No to display the next menu item. •...
  • Page 150: Activation Of A Binary Output As Pulse Output

    15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) 15.2 Activation of a binary output as pulse output PIOX S72* 15.2 Activation of a binary output as pulse output A pulse output is an integrating output which emits a pulse when the volume or the mass of the fluid which has passed the measuring point, reaches a certain value (pulse value).
  • Page 151: Activation Of A Frequency Output As Pulse Output

    15 Outputs (TF7-*721**-******-******a-*****-**; a = U, V, D) PIOX S72* 15.3 Activation of a frequency output as pulse output 15.3 Activation of a frequency output as pulse output Options\Channel A\Outputs\Frequency F1 • Select the list item Frequency F1 in the menu item Outputs. •...
  • Page 152: Outputs (Tf7-*72***-******-******A-*****-**; A = M, P, Q)

    16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) 16.1 Configuration of a digital output as binary output PIOX S72* Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) If the transmitter is equipped with outputs, they have to be configured. For the configuration of the analog output, see Options section 10.3.6.
  • Page 153 16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) PIOX S72* 16.1 Configuration of a digital output as binary output Options\Channel A\Outputs\Digital output B1\Enable B1 • Select Yes to change the settings for an already assigned output or to assign a new output. •...
  • Page 154 16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) 16.1 Configuration of a digital output as binary output PIOX S72* Output function test The function of the output can now be tested. • Connect a multimeter to the output. Options\Channel A\Outputs\...\B1 Test signal •...
  • Page 155: Configuration Of A Digital Output As Pulse Output

    16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) PIOX S72* 16.2 Configuration of a digital output as pulse output Tab. 16.5: Output function test – measuring range list item switching function test value description • binary output is energized Status OK Status OK (status value) •...
  • Page 156 16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) 16.2 Configuration of a digital output as pulse output PIOX S72* Tab. 16.6: Selection of the physical quantity source item list item output Pulse Pulse |V| pulse without considering the sign of the volumetric flow rate pulse for positive measured values of the Pulse +V...
  • Page 157 16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) PIOX S72* 16.2 Configuration of a digital output as pulse output 16.2.2 Pulse output by defining pulses per unit Options\Channel A\Outputs\Digital output B1\Source item\Pulse\Pulse output • Select Pulses per unit. • Press ENTER. Options\Channel A\Outputs\Digital output B1\...\Output range B1 •...
  • Page 158: Configuration Of A Digital Output As Frequency Output

    16 Outputs (TF7-*72***-******-******a-*****-**; a = M, P, Q) 16.3 Configuration of a digital output as frequency output PIOX S72* Tab. 16.7: Output function test – signal output mode test value description The entered test value has to be If the external measuring instrument displays the entered Pulses per unit within the output range.
  • Page 159: Inputs

    17 Inputs PIOX S72* 17.1 Configuration of an input Inputs The inputs are configured in the program branch Special functions and assigned to the individual measuring channels in the program branch Options. 17.1 Configuration of an input If the transmitter is equipped with inputs, they have to be configured.
  • Page 160 17 Inputs 17.1 Configuration of an input PIOX S72* Measuring range Special functions\Inputs\...\Start of meas. range • Enter the lowest expected measured value. The unit of measurement of the source item will be displayed. Start of meas. range is the physical quantity assigned to the lower limit of the input range (Input MIN). •...
  • Page 161 17 Inputs PIOX S72* 17.1 Configuration of an input 17.1.3 Definition of a switching condition If a transmitter function is to be performed by remote control, a switching condition has to be defined. Special functions\Inputs\...\Trigger value • Select Yes if a switching condition is to be defined. Select No to display the next menu item. •...
  • Page 162 17 Inputs 17.1 Configuration of an input PIOX S72* 17.1.4 Binary inputs The transmitter can be equipped with max. 4 binary inputs. Via the binary inputs, it is possible to remotely trigger some functions of the transmitter. Inversion of the input Special functions\Inputs\Binary Bx\Negate input •...
  • Page 163: Assignment Of An Input

    17 Inputs PIOX S72* 17.2 Assignment of an input Binary input • Connect the signal source to the input. Special functions\Inputs\...\I1 Test signal • Select Yes to test the input signal. Select No to display the next menu item. • Press ENTER. If the transmitter displays that a input signal is connected, the input functions correctly.
  • Page 164: Data Logger

    18 Data logger 18.1 Configuration of the data logger PIOX S72* Data logger The transmitter has a data logger which stores the measured values during the measurement. Special functions Data logger Notice! In order to store measured data, the data logger has Configuration to be configured.
  • Page 165 18 Data logger PIOX S72* 18.1 Configuration of the data logger display description The storing starts immediately. Immediately The storing starts in the next full 5 minutes. Full 5 minutes The storing starts in the next full 10 minutes. Full 10 minutes The storing starts in the next full 15 minutes.
  • Page 166 18 Data logger 18.1 Configuration of the data logger PIOX S72* Special functions\Data logger\Configuration\Start storing\Event-based\ Storage rate (trig.) • Select in the scroll list a storage rate for storing the measured values in case the event occurs. • Press ENTER. Special functions\Data logger\Configuration\Start storing\Event-based\ Buffer time ->П...
  • Page 167: Deletion Of The Data Logger

    18 Data logger PIOX S72* 18.2 Deletion of the data logger Storage mode Special functions\Data logger\Configuration\Storage mode • Select the menu item Configuration. • Press ENTER until the menu item Storage mode is displayed. • Press ENTER. • Select Sample to store the current measured value. •...
  • Page 168: Information Regarding The Data Logger

    18 Data logger 18.3 Information regarding the data logger PIOX S72* 18.3 Information regarding the data logger Special functions\Data logger\Data logger info • Select the menu item Data logger info. • Press ENTER. The following information regarding the data logger is displayed: display description data logger is activated/deactivated...
  • Page 169 18 Data logger PIOX S72* 18.5 Transmission settings Special functions\Data logger\Transmission settings\Delete spaces • Select On if the space characters are not to be transmitted. • Press ENTER. The file size will be considerably reduced (shorter transmission time). Special functions\Data logger\Transmission settings\Decimal marker •...
  • Page 170: Data Transmission

    19 Data transmission 19.1 Service interfaces PIOX S72* Data transmission The data is transmit via the service interfaces or the process interface (optional) of the transmitter. 19.1 Service interfaces By means of the program FluxDiagReader it is possible to transmit via service interfaces (USB, LAN) data to the PC. The following tasks can be carried out: •...
  • Page 171: Process Interface

    19 Data transmission PIOX S72* 19.2 Process interface 19.2 Process interface The transmitter can be equipped with a process interface (e.g., Profibus, Modbus). For the connection of the process interface to the transmitter see supplement to operating instruction. RS485 interface Special functions\Communication\RS485 •...
  • Page 172: Advanced Functions

    20 Advanced functions 20.1 Totalizers PIOX S72* Advanced functions 20.1 Totalizers The total volume or total mass of the fluid at the measuring point can be determined. There are 2 totalizers, one for the positive and the other for the negative flow direction. The unit of measurement used for totalizing corresponds to the volume or mass unit selected for the physical quantity.
  • Page 173 20 Advanced functions PIOX S72* 20.1 Totalizers decimal places max. value max. display < 10 ±9999999999 < 10 ±99999999.9 < 10 ±9999999.99 < 10 ±999999.999 < 10 ±99999.9999 Notice! The number of decimal places and the max. value of the totalizers only affect the display. 20.1.2 Detection of long measurement failures If there are no valid measured values during a long time interval, the totalizers remain unchanged.
  • Page 174: Calculation Channels

    20 Advanced functions 20.2 Calculation channels PIOX S72* 20.1.4 Totalizer behavior after the measurement is stopped It is possible to define the totalizer behavior after the measurement is stopped or after a reset of the transmitter. Special functions\Totalizers\Keep totalizers • Select the menu item Totalizers in the program branch Special functions. •...
  • Page 175 20 Advanced functions PIOX S72* 20.2 Calculation channels 20.2.2 Parameterization of a calculation channel Parameters\Channel Y • Select a calculation channel (here: Channel Y) in the program branch Parameters. • Press ENTER. The current calculation function is displayed. • Press ENTER. 20.2.2.1 Selection of the calculation type Tab.
  • Page 176 20 Advanced functions 20.2 Calculation channels PIOX S72* Parameters\Channel Y\Calculation type\-Upper limit • Select No limit if the calculation channel has to output all negative values without upper limit • Press ENTER. • Select Set to limit if the calculation channel has to output the limit when falling below the upper limit. •...
  • Page 177 20 Advanced functions PIOX S72* 20.2 Calculation channels 20.2.4 Measurement with calculation channels • Select the program branch Measurement. • Press ENTER. Measurement\Select channels • Activate the necessary channels. Calculation channels are activated or deactivated the same way as measuring channels.
  • Page 178 20 Advanced functions 20.2 Calculation channels PIOX S72* Tab. 20.3: Source item Extended diagnostics source item list item output percentage of physical channels, with valid measuring state Extended diagnostics Valid channels standard deviation of the sound speed σ(Sound speed) standard deviation of the flow velocity σ(Flow velocity) standard deviation of the signal amplification σ(Amplification)
  • Page 179: Diagnosis With The Help Of The Snap Function

    20 Advanced functions PIOX S72* 20.3 Diagnosis with the help of the snap function 20.3 Diagnosis with the help of the snap function 20.3.1 Configuration By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of measuring results or for diagnostic purposes.
  • Page 180: Modification Of The Limit For The Inner Pipe Diameter

    20 Advanced functions 20.4 Modification of the limit for the inner pipe diameter PIOX S72* 20.3.3 Information concerning snaps Special functions\Snap\Snap info • Select the menu item Snap info. • Press ENTER. The following information is displayed: display description number of stored snaps Stored snaps: number of snaps that can still be stored Snaps left:...
  • Page 181: Remote Functions

    20 Advanced functions PIOX S72* 20.5 Remote functions 20.5 Remote functions Remote functions can be triggered by triggerable analog inputs or event triggers. Options In order to define an input for a remote function, it has to be Channel x enabled in the menu item Special functions\Inputs.
  • Page 182 20 Advanced functions 20.5 Remote functions PIOX S72* Reset of measured values • Select the list item Reset meas. val. • Press ENTER. The measured value output simulates a reposing application for the duration of the signal. The actual measured flow velocity is ignored and the measured value is set to zero.
  • Page 183: Event Triggers

    20 Advanced functions PIOX S72* 20.6 Event triggers 20.6 Event triggers It is possible to configure max. 4 independent event triggers per channel: R1, R2, R3 and R4. Options The event triggers can be used, e.g., to: Channel x – output information about the running measurement –...
  • Page 184 20 Advanced functions 20.6 Event triggers PIOX S72* Tab. 20.4: Source item source item list item output fluid temperature Fluid properties Fluid temp. fluid pressure Fluid pressure fluid density Fluid density kinematic viscosity Kin. viscosity dynamic viscosity Dyn. viscosity standardized density Norm.
  • Page 185 20 Advanced functions PIOX S72* 20.6 Event triggers Definition of the switching condition Options\Event trigger\Rx Enable\Source item\...\Function • Select the switching condition. • Press ENTER. Definition of the holding behavior Options\Event trigger\Rx Enable\Source item\...\Type • Select the type of the holding behavior. •...
  • Page 186 20 Advanced functions 20.6 Event triggers PIOX S72* Definition of the switching delay Options\Event trigger\Rx Enable\Source item\...\Glitch interval • Enter a time interval after which the event trigger has to switch in case the event occurred. • Press ENTER. Options\Event trigger\Rx Enable\Source item\...\Failure delay •...
  • Page 187: Event Protocol

    20 Advanced functions PIOX S72* 20.7 Event protocol Tab. 20.6: Pictograms for the status display of the event triggers current state Function Type (switching condition) (holding behavior) deactivated MAX (x>limit) Non-hold (false state) activated MIN (x<limit) Hold (true state) Within range Hold for a while Out of range ERR (x=fail)
  • Page 188: Settings

    21 Settings 21.1 Dialogs and menus PIOX S72* Settings 21.1 Dialogs and menus Special functions\Dialogs/Menus • Select the menu item Dialogs/Menus in the program branch Special functions. • Press ENTER. Pipe circumference Special functions\Dialogs/Menus\Pipe circumference • Select the menu item Pipe circumference. •...
  • Page 189 21 Settings PIOX S72* 21.1 Dialogs and menus Error delay The error delay is the time after which an error value will be sent to an output if no valid measured values are available. Special functions\Dialogs/Menus\Error delay • Press ENTER until the menu item Error delay is displayed. •...
  • Page 190: Measuring Modes

    21 Settings 21.2 Measuring modes PIOX S72* Toggle time The toggle time indicates the time interval after which the transmitter toggles between the different channels during the measurement. The default value is 3 s. Special functions\Dialogs/Menus\Toggle time • Press ENTER until the menu item Toggle time is displayed. •...
  • Page 191: Measurement Settings

    21 Settings PIOX S72* 21.3 Measurement settings 21.3 Measurement settings Special functions\Measurement\Measurement settings • Select the menu item Measurement in the program branch Special functions. • Press ENTER. • Select the menu item Measurement settings. • Press ENTER. Multi-point calibration A multi-point calibration allows a very precise output of measuring results.
  • Page 192: Units Of Measurement

    21 Settings 21.4 Units of measurement PIOX S72* 21.4 Units of measurement It is possible to set the global units of measurement for length, temperature, pressure, sound speed, density and kinematic viscosity. Special functions\Units of measurement • Select the menu item Units of measurement. •...
  • Page 193: Working With Parameter Records

    21 Settings PIOX S72* 21.6 Working with parameter records 21.6 Working with parameter records 21.6.1 Introduction Parameter records are data sets that contain all information necessary to perform a certain measurement task: • pipe parameters • transducer parameters • fluid parameters •...
  • Page 194: Contrast Settings

    21 Settings 21.7 Contrast settings PIOX S72* 21.7 Contrast settings Special functions\System settings\Display contrast • Select the menu item System settings in the program branch Special functions. • Press ENTER. • Select the menu item Display contrast. • Press ENTER. The display contrast is adjusted with the following keys: increases the contrast reduces the contrast...
  • Page 195: Key Lock

    21 Settings PIOX S72* 21.9 Key lock 21.9 Key lock An ongoing measurement can be protected from an inadvertent intervention by means of a key lock. Definition of a key lock code • Select the menu item System settings in the program branch Special functions. •...
  • Page 196 21 Settings 21.9 Key lock PIOX S72* 2022-05-15, UMPIOX_S72xV1-9EN...
  • Page 197: Annex

    Annex PIOX S72* A Menu structure Annex Menu structure Program branches Parameters Measurement Options Special functions Channel x Select channels Channel x System settings Transducer Measuring point no. Measurement Inputs Outer diameter Sound path Units of measurement Outputs Pipe material Transducer distance Outputs Measurement...
  • Page 198 Annex A Menu structure PIOX S72* Analysis Special functions Measurement Measurement modes Enable analysis → Reference temp. → Reference pressure 2022-05-15, UMPIOX_S72xV1-9EN...
  • Page 199 Annex PIOX S72* A Menu structure Parameter input Parameters Channel x Transducer Connected transd. Select transducer Outer diameter Pipe material Pipe wall thickness Lining Lining 2 Coating Roughness Automatic Customized Fluid Fluid temp. Select analysis quant. → Mass fraction → Volume fraction →...
  • Page 200 Annex A Menu structure PIOX S72* Measurement settings Options Channel x Measurement Units of measurement Physical quantity Flow velocity Volumetric flow rate → Volumetric flow rate → Mass flow rate → Sound speed Volume total → Analysis quantity → Flow velocity Mass flow rate Unit of measurement Mass total...
  • Page 201 Annex PIOX S72* A Menu structure Data logger Special functions Data logger Configuration Delete meas. values Print meas. values Transmission settings Data logger info Logger channels Serial output channel Start storing Delete spaces → Activated: → Logger full: → Meas. val. series: →...
  • Page 202 Annex A Menu structure PIOX S72* Measurement start Parameters Measurement Select channels Measuring point no. Sound path Transducer distance For each Diagnostic values activated measuring channel Proposition of transducer distance Measurement Stop measurement Show measurement Show parameters Read-only mode 2022-05-15, UMPIOX_S72xV1-9EN...
  • Page 203 Annex PIOX S72* A Menu structure Field calibration Measurement Options Channel x Channel x Measuring point no. Field calibration c Sound path Auto calibration Auto activation Edit samples Clear samples Proposition of transducer distance Default Display of diagnostic values Transducer distance Measurement display Stop measurement UMPIOX_S72xV1-9EN, 2022-05-15...
  • Page 204 Annex A Menu structure PIOX S72* Configuration of outputs Options Special functions Channel x Current output Outputs Active Passive → Current → Voltage → Frequency → Binary Enable output Source item Sound speed Flow/analysis Totalizers Fluid properties Diagnostic values Miscellaneous Event trigger Pulse Pulse value...
  • Page 205 Annex PIOX S72* A Menu structure Configuration of inputs Special functions Inputs → Temperature → Current → Voltage → Binary Temperature Tx Enable Pt100/Pt1000 Temperature offset → Temperature Trigger value Function → Within range → MAX (x>limit) → Out of range →...
  • Page 206 Annex A Menu structure PIOX S72* Configuration of inputs Special functions Current Inputs Ix Enable → Temperature → Current → Voltage → Binary Source item Customized Current → Temperature → Pressure → Density Input name → Kin. viscosity → Dyn. viscosity Unit of measurement Decimal places Input range...
  • Page 207 Annex PIOX S72* A Menu structure Configuration of inputs Special functions Voltage Inputs Ux Enable → Temperature → Current → Voltage → Binary Source item Customized Voltage → Temperature → Pressure → Density Input name → Kin. viscosity → Dyn. viscosity Unit of measurement Decimal places Binary...
  • Page 208 Annex A Menu structure PIOX S72* Special settings Options Channel x Pipe signal detection Special settings Default Customized Cut-off flow LWT pipe wall calibr. Default Customized Default Flow velocity limit Linear calibration Default Customized Default Pig detection Profile correction Global settings Default Turbulence mode Weighting factor...
  • Page 209 Annex PIOX S72* A Menu structure Calculation channels Parameters Options Channel x Channel x Calculation Measurement Units of measurement Calculation type Physical quantity Flow velocity Average (all chan. OK) Average (1 chan. OK) Special Volumetric flow rate (AND) (OR) → Volumetric flow rate →...
  • Page 210 Annex A Menu structure PIOX S72* Results – overview Trigger Input quantities Physical quantities Special functions Options → Temperature → Flow quantities → Current → Sound speed → Voltage → Totalizers → Binary → Fluid properties → Diagnostic values Condition Event trigger →...
  • Page 211 Annex PIOX S72* A Menu structure Definition of event triggers Options Channel x Event trigger Select Rx Rx Enable Source item Sound speed Flow/analysis Totalizers Fluid properties Diagnostic values Miscellaneous Function → MAX (x>limit) → MIN (x<limit) → Within range →...
  • Page 212 Annex A Menu structure PIOX S72* Remote functions Trigger Special functions Options Inputs Channel x Condition Event trigger Temperature Current Voltage Binary Enable Enable Enable Enable Select Rx Pt100/Pt1000 Source item Source item Negate input Enable Rx Source item → Yes →...
  • Page 213 Annex PIOX S72* A Menu structure Event-based storing of measured values Trigger Special functions Options Inputs Channel x Condition Event trigger Temperature Current Voltage Binary Enable Enable Enable Enable Select Rx Pt100/Pt1000 Source item Source item Negate input Enable Rx Source item →...
  • Page 214 Annex A Menu structure PIOX S72* Example 1 The diagnostic values are to be checked. If the SCNR is < 20 dB, a snap has to be triggered. Trigger: SCNR ˂ 20 dB Condition: R1 with SCNR < 20 dB Action: trigger a snap Options...
  • Page 215 Annex PIOX S72* A Menu structure Example 2 The storage rate of all measured and diagnostic values of a certain temperature range is to be changed. The normal storage rate of all measured and diagnostic values is 1 h. If the temperature is outside the defined range of 20...40 °C, the storage rate should be 1 min.
  • Page 216: B Units Of Measurement

    Annex B Units of measurement PIOX S72* Units of measurement Length/roughness unit of measurement description millimeter inch Temperature unit of measurement description degree Celsius °C degree Fahrenheit °F Pressure unit of measurement description bar (absolute) bar(a) bar (relative) bar(g) pound per square inch (absolute) psi(a) pound per square inch (relative) psi(g)
  • Page 217 Annex PIOX S72* B Units of measurement Volumetric flow rate unit of measurement description default volume (totalized) cubic meter per day m3/d cubic meter per hour m3/h cubic meter per minute m3/min cubic meter per second m3/s cubic kilometer per hour km3/h milliliter per minute ml/min...
  • Page 218 Annex B Units of measurement PIOX S72* unit of measurement description default volume (totalized) gallon per day Igpd (Imp-gal/d) Igal gallon per hour Igph (Imp-gal/h) Igal gallon per minute Igpm (Imp-gal/m) Igal gallon per second Igps (Imp-gal/s) Igal imperial kilogallon per minute IKGM (Imp-Kgal/m) million imperial gallons per day IMGD (Imp-Mgal/d)
  • Page 219: C Reference

    The following tables provide assistance for the user. The accuracy of the data depends on the composition, temperature and processing of the material. FLEXIM does not assume liability for any inaccuracies. Sound speed of selected pipe and lining materials at 20 °C The values of some of these materials are stored in the internal database of the transmitter.
  • Page 220 Annex C Reference PIOX S72* Typical roughness values of pipes The values are based on experience and measurements. material absolute roughness [mm] drawn pipes of non-ferrous metal, glass, plastics and light metal 0…0.0015 drawn steel pipes 0.01…0.05 fine-planed, polished surface max.
  • Page 221 Annex PIOX S72* C Reference Typical properties of selected fluids at 20 °C C.3.1 Mixtures with non-changing composition fluid explanation sound density kinematic range of ap- speed [m/s] [g/cm viscosity plication [°C] liquid water 1482 0.999 0…350 Water liquefied under pressure -180…+97 Propane liquefied under pressure,...
  • Page 222 Annex C Reference PIOX S72* C.3.2 Mixtures with changing composition fluid explanation parameter- sound density kinematic range of ization speed [g/cm viscosity application [m/s] ethylene glycol proportion 1482…1710 999…1132 1…23 -30…+150 °C Glycol/H2O of glycol 0…100 % hydrocarbon API gravity 530…1800 440…1130 1…>...
  • Page 223 Annex PIOX S72* C Reference Properties of water at 1 bar and at saturation pressure fluid temperature fluid pressure sound speed density specific heat capacity [°C] [bar] [m/s] [kg/m³] [kJ/kg/K 1.013 1402.9 999.8 4.219 1.013 1447.3 999.7 4.195 1.013 1482.3 998.2 4.184 1.013...
  • Page 224: D Conformity Declarations

    Annex PIOX S72* D Conformity declarations Conformity declarations For EU declaration of conformity according to ATEX directive, see document SIFLUXUS. UMPIOX_S72xV1-9EN, 2022-05-15...

Table of Contents