Image Brightness Setting - µ-Dimension AutoGet Wifi User Manual

Intelligent fiber endface microscope
Table of Contents

Advertisement

3.3.3 Image brightness setting

SF image brightness setting.
If the brightness of the SF image is not suitable and you
need to adjust it, follow the steps below:
1. Select SF task
2. Go to Settings - Devices, select SF, and then adjust the
image brightness. The brightness of the SF image is
recommended to be adjusted between 185 and 195.
MF image brightness setting.
If the brightness of the MF image is not suitable and you
need to adjust it, follow the steps below:
1. Select MF measurement task to return to the main
interface of the software.
2.Go to Settings - Devices, select MF, and then adjust the
image brightness. The brightness of the MF image is
recommended to be adjusted between 75 and 85.
AutoGetWifi Intelligent Fiber Endface Microscope
25
Figure3-16
Figure3-17