5. Repeat the operations specified in items 1 to 4 for 10 and 60 mm thick test
samples.
The calibration test results are considered positive if the obtained values of the
Device measurement error for all test samples do not exceed ± 0.04 mm.
MAINTENANCE
Transducer Replacement
When the plastic coating of the transducer is worn out, replace the transducer.
Battery Replacement
The Device is equipped with a removable battery; the user can replace the
battery for a new one, if necessary. Replace the battery once per three years.
+7(812) 385-54-28
info@oktanta-ndt.ru
EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
Replace the transducer when the Device is off.
WARNING!