Thickness Measurements Using The Device - OKTANTA EM1401 Operation Manual

Thickness gauge
Table of Contents

Advertisement

from -20 to
EMT14014T
+750
from -5 to
EDC10P5F3
+50
from -5 to
EDC5P7.2FS10
+50
from -5 to
EDC5P10FS15
+50
Note that, when using the EMT14014T transducer, it cannot be heated to the
temperatures exceeding 300 ºС. Therefore, while operating with the objects heated to
the temperatures exceeding 300 ºС, it is necessary to observe time intervals in
accordance with the table below:
Temperature range, ºС
0 – 300
300 – 700

Thickness Measurements using the Device

To measure the thickness, do as follows:
1.
Select the transducer to operate with the thickness gauge as described in
the previous item.
2.
If necessary, clean the surface and apply contact liquid (when using the
piezoelectric transducers).
+7(812) 385-54-28
info@oktanta-ndt.ru
EM1401 / EM1401 UT EMA Thickness Gauge. Operation Manual.
from 2 to
from 0 to
120 mm
3
from 0.5
-
to 20
from 1 to
-
150
from 2 to
-
300
0 – 5 (time interval between measurements:15
The transducer has a steel housing and
special heat-resistant protector and is
designed for operation with the objects
heated up to 750 C.
The transducer operates at 10 MHz and
is designed for thickness measurements
of the thin objects and small defects
detection.
The transducer operates at 5 MHz and is
designed for thickness measurements of
the medium thick objects and suitable
for most of applications.
The transducer operates at 5 MHz and is
designed for thickness measurements of
the thick objects and of the objects with
poor surface quality.
EMT14014T transducer measurement modes
Measurement duration, seconds
continuously
seconds)