Commissioning
For example:
If the design flowrate is 6 T/H, then:
Therefore the test weight should be between 162 and 216 g.
If an available test weight of 200 g is used, then:
Perform the zero calibration as described in the Calibration section of the integrator manual.
After the completion of the zero calibration, a span calibration, as described in the integrator
instruction manual, can be performed with the test weight(s) applied.