Installation
1. Set the flowmeter housing in place ensuring that the flowguide and outlet gaskets are
correctly located (gaskets supplied by customer).
2. The housing must be vertical within 1° in two planes 90° apart.
3. Bolt the flowguide and outlet flanges to the process piping.
4. Remove the shipping stop
Note
The hole in the enclosure for the shipping stop should be plugged with a bolt or cap to
ensure a dust-tight seal.
5. Connect the unit to the integrator
SITRANS WF100
Operating Instructions, 08/2021, A5E51108132-AA
Infeed (customer supplied)
Mounting hardware
Gasket
SITRANS WF100
Discharge (customer supplied)
Calibration pulley
Installing/Mounting
21