Dc Voltage Specifications - Fluke 5520A Operator's Manual

Multi-product calibrator
Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

1-14. DC Voltage Specifications

Absolute Uncertainty, tcal ± 5 °C
Range
15 + 1
0 to 329.9999 mV
9 + 2
0 to 3.299999 V
10 + 20
0 to 32.99999 V
15 + 150
30 V to 329.9999 V
15 + 1500
100 V to 1000.000 V
300 + 350
0 to 329.999 mV
300 + 350
0.33 V to 3.29999 V
300 + 350
3.3 V to 7 V
TC Simulate and Measure in Linear 10 µV/°C and 1 mV/°C modes [3]
40 + 3
0 to 329.999 mV
Remote sensing is not provided. Output resistance is < 5 mΩ for outputs ≥ 0.33 V. The AUX output has an output
[1]
resistance of < 1 Ω. TC simulation has an output impedance of
[2]
Two channels of dc voltage output are provided.
[3]
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
Range
0 to 329.9999 mV
0 to 3.299999 V
0 to 32.99999 V
30 to 329.9999 V
100 to 1000.000 V
0 to 329.999 mV
0.33 V to 3.29.999 V
3.3 V to 7 V
[1]
Two channels of dc voltage output are provided.
± (ppm of output + µV)
90 days
1 year
20 + 1
11 + 2
12 + 20
18 + 150
18 + 1500
Auxiliary Output (dual output mode only) [2]
400 + 350
400 + 350
400 + 350
50 + 3
Bandwidth 0.1 Hz to 10 Hz p-p
± (ppm output + floor)
0 + 1 µV
0 + 10 µV
0 + 100 µV
10 + 1 mV
10 + 5 mV
Auxiliary Output (dual output mode only) [1]
0 + 5 µV
0 + 20 µV
0 + 100 µV
Introduction and Specifications
Stability
24 hours, ± 1 °C
Resolution
± (ppm output + µV)
3 + 1
0.1
2 + 1.5
1
2 + 15
10
2.5 + 100
100
3 + 300
1000
30 + 100
1
30 + 100
10
30 + 100
100
5 + 2
0.1
± 1
10
.
Noise
Bandwidth 10 Hz to 10 kHz rms
6 µV
60 µV
600 µV
20 mV
20 mV
20 µV
200 µV
1000 µV
Specifications
Max
Burden
µV
[1]
50 Ω
10 mA
10 mA
5 mA
5 mA
5 mA
5 mA
5 mA
10 Ω
1
1-11

Advertisement

Table of Contents
loading

Table of Contents