66 | DR Full Leg Full Spine | Problem solving

Anatomical stitching is not optimal

Table 9: Problem: Anatomical stitching is not optimal
Details
Cause
Brief Solution
0179E EN 20210708 1527
The anatomic information in the partial images cannot
be automatically fully aligned. If the result of automatic
stitching based on the alignment of the anatomical in-
formation in the image is suspicious because of possible
movement of the patient during the examination, this
icon is displayed in the Stitching pane:
The patient has changed position during the examina-
tion.
Manually adjust the DR Full Leg Full Spine image.
If the partial images cannot be manually adjusted, click
Cancel in the Stitching pane. No DR Full Leg Full Spine
image is available.