Prepare the Full Leg Full Spine configuration

In the examination room, position the X-Ray system and the DX Full Leg Full
Spine Stand:
1. To use a portable DR Detector, insert it in the radiographic wall stand DR
bucky.
Depending on the configuration, the DR Detector can be inserted in
portrait or landscape position.
2. On the X-Ray system control panel or remote control, press and hold the
button that moves the X-ray system to the automatic position.
The X-Ray system moves to the default position for preparing the DR Full
Leg Full Spine examination. The default position of the X-ray tube, is such
that the touch screen console is easy to reach.
3. Position and fixate the DX Full Leg Full Spine Stand.
Before moving the stand from its parking position, release the brakes on
the four wheels.
When moving the stand over a longer distance, turn the stand to its lateral
direction so it will not block your sight.
Two floor mount locks are located at the side of the DX Full Leg Full Spine
Stand. The exact position depends on the configuration. Position the locks
right above the indicated position on the floor. Push the upper lever to
fixate the lock.
WARNING:
Wrong image orientation will cause the stitching to fail.
Follow the instructions in the user manual of the DR
detector for positioning the detector in the bucky.
WARNING:
Fixate both floor mount locks before positioning the patient.
WARNING:
The housings of the floor mount locks can cause a tripping
hazard.
When the DX Full Leg Full Spine Stand is not used, position
it in such a way that tripping over the floor mount locks is
avoided.
DR Full Leg Full Spine | Getting started | 29