AGFA DR Full Leg Full Spine User Manual page 33

Table of Contents

Advertisement

ROI Top
156
ROI Bottom
56
Positioning
Figure 9: Values for region of interest on DX-D 300
• Top: read the top of the region of interest on the vertical ruler of the
DX Full Leg Full Spine Stand. This is the start position.
• Bottom: read the bottom of the region of interest on the vertical ruler
of the DX Full Leg Full Spine Stand. This is the end position.
• Distance/OID: if the image is used for making length measurements in
NX or on true size printed images, read the distance between the
stitching grid of the DX Full Leg Full Spine Stand and the plane in
which measurements are to be made from the horizontal ruler on the
handle of the DX Full Leg Full Spine Stand. This distance is used for
calibrating distance measurements on the image in NX. If no
calibration is required, enter zero.
3. Set the X-Ray tube angle to 0 degrees.
4. On the collimator, switch on the light localizer.
If required, collimate in transversal direction.
Related Links
Making measurements
STITCHING
OID
0
Set
Cancel
Programs
Generator
Note: The system imposes limits to the bottom and the top of
the region of interest, depending on the X-Ray system
installation.
Instruction: For automatic stitching, the width of the
collimated area must be more than 15 cm.
on page 63
DR Full Leg Full Spine | Getting started | 33
0179E EN 20210708 1527