DR Full Leg Full Spine | Introduction to the Full Leg Full Spine application | 23

Safety Directions

WARNING:
Safety is only guaranteed when an Agfa certified field service
engineer has installed the product.
WARNING:
The following actions may lead to serious risk of injury and
damage to the equipment as well as making the warranty void:
Changes, additions or maintenance to the Agfa products carried
out by persons without appropriate qualifications and training.
Using unapproved spare parts
WARNING:
Do not move or position the DX Full Leg Full Spine Stand on
ramps with an inclination exceeding 10°. Moving or positioning
the DX Full Leg Full Spine Stand on ramps with higher
inclination can cause serious damage to the equipment and
represents a danger for the user and the patient.
WARNING:
Handle the FLFS Horizontal Overlay with care to avoid damage.
WARNING:
Do not use the FLFS Horizontal Overlay if it is damaged to avoid
injury.
WARNING:
Monitor the system movements with special care. Avoid any
impact of the system with walls, furniture or other elements in
the room that may cause damage to the equipment.
WARNING:
The patient can fall from the DX Full Leg Full Spine Stand. Make
the patient aware of the step and provide assistance to climb or
descend from the DX Full Leg Full Spine Stand.
CAUTION:
All Agfa medical products must be used by trained and qualified
personnel.
0179E EN 20210708 1527