Internal Pump Menu (Option); Valve Menu - ACOEM Met One Instruments GAS-1060 User Manual

Direct nitrogen dioxide analyzer us epa designated equivalent method
Table of Contents

Advertisement

Serinus 60 User Manual 3.0
Meas. Count
Sig. Count 1
Sig. Count 2
Sig. Count 3
Sig. Count 4
LED Drive
3.4.15

Internal Pump Menu (Option)

Main Menu  Service Menu  Diagnostics Menu  Internal Pump Menu
Flow Control
Internal Pump
Coarse
Fine
Sample Flow
Internal Pump
3.4.16

Valve Menu

Main Menu  Service Menu  Diagnostics Menu  Valve Menu
The Valve Menu allows the user to observe the instrument controlled switching of the valves. If the
valve is On it means the valve is energised. When a three way valve is in the On state it will now be in
the NC (normally closed) position as shown in the plumbing schematic. When the valve sequencing is
disabled the user has the ability to turn the valve Off and On manually. It is recommended that the
valve menu be used by a trained technician following the plumbing schematic (refer to Section 9.5).
Note: When interpreting the information below regarding the flow path through the valve note that
(NC = Normally Closed), (NO = Normally Open) and (C = common).
Page 50
0 - 15
This value increases by 2 every 5 seconds.
0 - 16777215
A raw intensity count of the response during the first 90° of the signal
phase.
0 - 16777215
A raw intensity count of the response during the second 90° of the
signal phase.
0 - 16777215
A raw intensity count of the response during the third 90° of the
signal phase.
0 - 16777215
A raw intensity count of the response during the final 90° of the signal
phase.
0-255
Controls the current used to drive the LED thus varying the intensity.
Set to MANUAL to disable the automatic flow control. AUTO allows
the flow PID to modify the pump coarse and fine settings. START will
transition to AUTO after one second.
This field allows the internal pump to be turned on or off. This field
is only editable when the Flow Control field is set to MANUAL.
Internal pump speed control (Coarse). This field is only editable
when the Flow Control field is set to MANUAL.
Internal Pump speed control (Fine). This field is only editable when
the Flow Control field is set to MANUAL.
Current gas flow. This is only accurate when reading close to the
flow calibration point.
The current pressure reading of the flow block which is down stream
of the orifice.