Connecting An Endoscope - Pentax EPK-i Owner's Manual

Hide thumbs Also See for EPK-i:
Table of Contents

Advertisement

Connecting an Endoscope

Connect a scope to the processor. Connect the water bottle and suction unit to the scope. For details on the scope, refer to the
instruction manual for the scope.
1
Align the scope locking lever of the processor to the
OPEN position.
2
Insert the electrical connector and light guide of the
scope into the corresponding scope electrical
connector
and light guide attachment of the
1
processor and move the scope locking lever in the
direction of
until it clicks into the LOCK position.
2
3
Pour sterile water into the water bottle until the bottle is about two thirds full and then securely
tighten the cap.
4
Insert the water bottle nozzle into the bottle connector of the processor.
5
Align the A/W and DRAIN switch lever of the water
bottle to the A/W (air/water supply) position.
•• Insert the air/water supply plug into the holder.
2
1
Light Guide
A / W
Preparation
Scope
Locking
Lever
2
Electrical
Connector
DRAIN
15