Installation - Pentax EPK-i Owner's Manual

Hide thumbs Also See for EPK-i:
Table of Contents

Advertisement

Installation

Installing the Processor in a Cart
Install the processor and peripheral devices while referring to the installation examples in “"System Configuration”" (P.11). Install
the processor and peripheral devices in a cart. Make sure the cart is on a stable and level surface.
Do not install the processor in any of the following locations.
- Where the processor is likely to be exposed to water.
- Where flammable or explosive gas is present.
- In hot and humid locations.
2
- Where the processor will be exposed to direct sunlight.
Install the processor in a location where the ventilation grids will not be blocked.
Install the processor in a location where dust will not enter inside. Wipe off any dust on the processor. Excessive
amounts of dust accumulating inside the unit may cause the processor to malfunction, emit smoke, or catch fire.
Connecting an Isolation Transformer
Connect the processor and peripheral devices to an isolation transformer. To reduce the possibility of an electric shock, be sure
to supply power to the processor and peripheral devices from the SAT-1300 isolation transformer specified by PENTAX or
another medical isolation transformer.
Isolation Transformer
1
Connect the power cord of the isolation transformer to a three-prong power outlet that meets the
power rating indicated on the rating plate.
2
Use the AC power cord to connect the power input socket on the rear panel of the processor to
the isolation transformer.
•• Make sure the processor is turned off beforehand.
3
Connect the peripheral devices to the isolation transformer.
•• Make sure the peripheral devices are turned off prior to installation.
•• For details on a peripheral device, refer to the instruction manual for the peripheral device.
Do not connect any electrical devices other than the processor and peripheral devices of the processor to the isolation
transformer.
Make sure the voltage, current, and power consumption of the processor and peripheral devices do not exceed the
maximum ratings indicated on the isolation transformer.
Be sure to securely connect the power cord of the isolation transformer to a three-prong power outlet.
12
Preparation
Peripheral Device
Processor
AC Power Cord

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents