  • Page 1 Cole-Parmer® HO-400 Series Homogenizer User Manual...
  • Page 2 Data herein has been verified and validated. It is believed adequate for the intended use of the instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, confirmation of the validity and suitability should be obtained;...
  • Page 3 Overview Digital Display Power Cord Port Speed Control On/Off Switch Timer Control Cross Rod Port (stand mounting) Run Button Generator Probe Coupling Stop Button Cross Rod Clamp The HO-400 Homogenizer consists of the following: The HO-400 consists of the following: Description Quantity Motor Drive Unit...
  • Page 4 Stand Mounting 1” Cross Rod Port Cross Rod Clamp Thumb Screw Assemble the stand (sold separately). Insert the cross rod into the cross rod port at the back of the unit and tighten thumb screw by turning clockwise until tight. Stand Part Loosen the second knob on the post No.04727-89...
  • Page 5 Accessories Accessories Generator Probes Generator Probes 1. Insert Generator 2. Quarter turn 3. Ready to use Probe into coupling clockwise to lock *always submerge probe in liquid before starting ¼ turn tTurn Generator Coupling *Never run dry Generator Probes Generator Probes Size Processing Range Type...
  • Page 6 Operation Speed Time 0:15 (back of unit) 1. D 4. Timer Control . Digital Display 7. Stop Button 5. RPM 8. On/Off switch 2. Timer 6. Run Button 3. Speed Control General Operation - Plug in the power cord and move the On/Off switch on back of the unit to the On position.
  • Page 7 Troubleshooting Error Code Definition Cause Possible Solution Unplug and let stand for at least 1 hour at Motor has overheated room temperature, then attempt to restart. Replace brushes using Motor brushes could be worn 20-110-05R or 20-220-05R down from heavy use. depending on motor voltage rating.