Thermo Scientific BF51728 Installation And Operational Manual page 45

Lindberg/blue m moldatherm 1100 c box furnace
Table of Contents

Advertisement

| Main Controller: 1x8 & 5x16 Segment Programmable
Chapter 6
|
6-44
330985H01
End Type parameter
The action which occurs at the end of program or in reset depends on the configuration of
the 'END.T' parameter. The 'END.T' can be:
OFF : The heating is turned OFF,
dwEll : Controls at last program setpoint,
SP2 : Controls at setpoint 2 (When the timer completes the target setpoint will switch to
setpoint 2. The setpoint 2 may be a lower or a higher temperature.),
rES : Reset on completion and reverts to SP1 or SP2, based on the Setpoint selection.
SCROLL
through parameters in level 2 and set the required 'END.T' by pressing
or
.
OR
To Configure the Programmer
1. Enter level 2 to configure the timer as a programmer. Press SCROLL
scroll through the list of parameters until 'TM.CFG' is displayed. Now, press
DOWN
or UP
2. To set the resolution, press SCROLL
or UP
button to set 'Hour or 'min' (In this example, the ramp rate and dwell
period are set in hours).
OR
button to select 'PROG'.
button to select 'TM.RES'. Press DOWN
OR
button to
1100°C Box Furnace