Thermo Scientific BF51728 Installation And Operational Manual page 26

Lindberg/blue m moldatherm 1100 c box furnace
Table of Contents

Advertisement

1100°C Box Furnace
4. After each adjustment recheck the door seal.
The door seal has been adjusted properly if there is no heat loss when operating the
furnace up to 500°C.
WARNING : When installing, maintaining, or removing the fiberglass insulation, the
following precautions will minimize airborne dust and fiber:
Keep personnel not involved in the installation out of the area.
Use a good vacuum to clean area and equipment. Use a dust suppressant if
sweeping is necessary. Do not use compressed air.
Use a disposable mask suitable for nuisance dust.
Wear long sleeve clothing, gloves, hat, and eye protection to minimize skin and
eye contact. Do not wear contact lenses.
Thoroughly wash self after work is complete.
Launder work clothing separate from other clothes and thoroughly clean
laundering equipment after use. If clothing contains a large amount of dust and/or
fiber, dispose of rather than clean.
Promptly place used ceramic fiber parts and dust in plastic bags and dispose of
properly.
Start-Up |
Chapter 5
330985H01 | 5-25