Delayed Switch On Timer - Thermo Scientific BF51728 Installation And Operational Manual

Lindberg/blue m moldatherm 1100 c box furnace
Table of Contents

Advertisement

| Main Controller: 1x8 & 5x16 Segment Programmable
Chapter 6
|
6-42
330985H01
Soft Start Power Limit
The soft start function limits the power delivered to the heater until it has warmed up. The
SS.PWR is the power limit applied until the PV reaches the SS.SP or the timer has
elapsed.

Delayed Switch On Timer

The timer is used to switch on the controller output power after a fixed length of time. It
could be used to turn on a process at a particular time.
It is initiated by any of the following:
Switching on power;
Momentarily pressing
Setting the parameter T.STAT to run;
A command through serial communications;
A logic input suitably configured.
When the timer status = run, the control output is off.
When the timer status = reset, the control output is controlling.
and
together;
1100°C Box Furnace

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bf51748Bf51766Bf51794Bf51828Bf51848Bf51866 ... Show all

Table of Contents