Chapter 4 Installation; Lifting And Carrying; Transport; Location - Thermo Scientific BF51728 Installation And Operational Manual

Lindberg/blue m moldatherm 1100 c box furnace
Table of Contents

Advertisement

1100°C Box Furnace
Installation
CAUTION : Be sure ambient temperature does not exceed 40°C (104°F). Ambient
above this level may result in damage to the controller. The recommended ambient
temperature is 17°C to 27°C (62.6°F to 80.6°F).
CAUTION : Allow at least 3" (7.62 cm) of space between the furnace, at least
12" (30.48 cm) above the furnace and any combustible surface. This permits the heat
from the furnace case to escape so as not to create a possible fire hazard.
WARNING : To avoid electrical shock, this furnace must be installed by a competent
electrician who ensures compatibility among furnace specification, power source and
ground code requirements.

Lifting and Carrying

CAUTION : Heavy loads. Lift with care!
1. To avoid injury through physical strain, such as strain trauma and slipped discs, do
not attempt to lift the furnace alone.
2. To avoid injury through dropped loads, wear Personal Protective Equipment (PPE)
per local protocols, such as safety shoes, when lifting the furnace.
3. To avoid crushing your fingers or hands (particularly in a closing door) or damaging
the furnace, do not use any other lift points than the bottom sides of the furnace.

Transport

1. For transport, do not lift the device using the doors or components attached to the
device (e.g. control box on rear panel) as lift points.
2. Lift at the bottom sides of the furnace with NIOSH rated straps and/or using
appropriate number of personnel per local safety policies and regulations.

Location

Install the furnace in a level area free from vibration. To permit proper air flow, leave at
least 3" (7.62 cm) of space on all sides of the unit and 12" (30.48 cm) above the unit.

Wiring

For detailed wiring information, refer to:
Figure 10 "Wiring Diagram (BF51728, BF51828)",
Figure 11 "Wiring Diagram (BF51748, BF51766, BF51848 & BF51866)" or
Figure 12 "Wiring Diagram (BF51794, BF51894)".
WARNING : Detachable Cordsets cannot be replaced with inadequately rated cords.