Installation

Section 5
Space allotment for a still should include a 6" clearance at the sides and
top to allow for cover removal and air circulation. Unit must be placed
within 4 feet of water supply and should not be mounted over electrical
equipment to avoid shock hazard.
Warning Do not mount your MEGA-PURE 12A Water Still directly over
equipment that requires electrical service. Routine maintenance of this unit
may involve water spillage and subsequent electrical shock hazard if
improperly located.
s
Warning Wall composition and construction, as well as fastener type, must
be considered when mounting this unit. The mounting surface and
fasteners selected must be capable of supporting a minimum of 150 lbs.;
inadequate support and/or fasteners may result in damage to mounting
surface and/or equipment. If unsure of mounting surface composition,
condition and construction, or correct fasteners, consult your building
maintenance group or contractor.
s

Wall Mount

The MP-12A Water Still has slotted holes for wall mounting. To support
the weight of this unit, use ¼" x 1¼" lag bolts for mounting. An optional
wall mounting bracket is available for the Automatic Collection System.

Bench Mounting

No special consideration for a "still only" setup. When used with the ACS,
the Water Still must be placed on a 6" platform to allow for gravity feed of
distilled water to the ACS.
Thermo Scientific
12A Water Still
5-1