Error Codes

Background
Code
Refers to
Display flashes
Cavity
set point
temperature
higher than set
point
Co m bi n a ti o n W ar m er
Op e r at or ' s Man u a l
This section is provided for the assistance of qualified and trained service
technicians only and is not intended for use by untrained or unauthorized service
personnel. Failure to observe this precaution may void the warranty.
NOTE:
If the warmer is not operating properly, check the following
before calling an authorized service agent:
Verify that the power to the warmer is on.
If applicable, ensure the female end of plug is securely seated in the
warmer and that the male end of plug is in an appropriate,
functioning outlet.
If applicable, examine the fuses. Replace the fuses, refer to topic How
to Replace a Fuse.
If applicable, examine the high limit manual reset button. If the high
limit manual reset button is tripped, reset the warmer, refer to topic
How to Manually Reset the Warmer.
NOTE:
All non-critical codes can be cleared using the standby button.
Critical errors (marked with a *) can only be cleared by setting the
power switch at the rear of the warmer to the off (O) position and
allowing the warmer to cool.
Do not attempt to repair or service the warmer beyond this
NOTICE
point. Contact the manufacturer for the nearest authorized
service agent. Repairs made by any other service agent
without prior authorization by the manufacturer will void the
warranty.
Action required
Cavity temperature is higher than the set point temperature.
◾Allow cavity to cool to set point temperature.
MN -4 7 42 4
Re v 1
4 / 2 1
T
ROUBLESHOOTING
49